Abstract 


Background

Laboratory testing to identify the molecular basis of serologic weak D phenotypes is recommended to determine whether a pregnant woman or potential transfusion recipient should be managed as RhD-positive or RhD-negative. The variation in D antigen expression on RBCs, different potencies of anti-D typing reagents, lack of standardized test methods, and the subjectivity of interpreting agglutination reactions complicate the detection of D variants. We evaluated the correlation of agglutination scores by an automated immunoassay analyzer with D antigen densities determined by flow cytometry, and D variant types identified by molecular analysis.

Materials and methods

We selected 273 blood donor samples with agglutination scores of less than 92 (4+), measured by an automated analyzer (NEO®, Immucor, Norcross, GA, USA). D antigen densities were measured by flow cytometry for 89 samples. Samples were classified as molecularly-determined weak D or partial D variants by multiplex PCR, PCR RFLP and DNA sequencing.

Results

All samples with a D antigen density ≥15% had an agglutination score >80 (4+). Agglutination scores for weak D types varied from 10 to 90. Agglutination scores for partial D antigens were graded with scores varying from 60 to 99. D antigen densities varied from 0.55% to 10.67% for weak Ds and 4.1% to 30.5% for partial Ds.

Discussion

Our results showed that score values follow a pattern among D variants that could be related to antigen density and to the RhD variant classification.