Abstract
Background
While common, studies assessing outcomes of failed metal-on-metal (MoM) resurfacings converted to total hip arthroplasties (THAs) are limited. We determined the outcomes following revision THA of aseptic MoM hip resurfacings.Methods
Between 2000 and 2019, we identified 52 revision THAs for failed MoM hip resurfacings through our total joint registry. Mean age was 55 years, 42% were women, and mean body mass index was 28 kg/m2. Adverse local tissue response led to THA in 67% of the cases. The most common revision articulation was metal-on-cross-linked or ceramic-on-cross-linked polyethylene (71%). Median head size was 36 mm. The acetabular component was retained in 21% and 28% used dual-mobility constructs. The mean follow-up was 6 years (range, 2-12 years).Results
The 5-year survivorships free of any re-revision or reoperation were 89% and 85%, respectively. The primary cause of re-revision (6) was dislocation (4). The 5-year cumulative probability of dislocation was 19% and was 13% in those patients treated with dual-mobility constructs versus 22% in those treated with standard articulations (P = .58). No dislocations occurred in THAs with retained acetabular components and dual-mobility constructs versus a 5-year cumulative probability of dislocation of 25% in those with revised acetabular components and standard articulations (P = .24).Conclusion
Revision THAs for aseptically failed MoM hip resurfacings yielded a 5-year survivorship free of re-revision of 89%. The main reason for failure was dislocation, which was reduced, but not statistically significantly, when a dual-mobility construct was used, especially if the acetabular component was retained.Level of evidence
IV.Citations & impact
Impact metrics
Citations of article over time
Alternative metrics
Discover the attention surrounding your research
https://www.altmetric.com/details/139169628
Article citations
Revision of metal-on-metal hip replacements with dual-mobility bearings and acetabular component retention.
Bone Jt Open, 5(6):514-523, 24 Jun 2024
Cited by: 1 article | PMID: 38910515 | PMCID: PMC11194627
Profiling the Immune Response to Periprosthetic Joint Infection and Non-Infectious Arthroplasty Failure.
Antibiotics (Basel), 12(2):296, 01 Feb 2023
Cited by: 6 articles | PMID: 36830206 | PMCID: PMC9951934
Review Free full text in Europe PMC
Metal-on-Metal Hips: Ten-Year Clinical and Radiographic Outcomes of the ADEPT Metal-on-Metal Hip Resurfacing and Modular Total Hip Arthroplasty.
J Clin Med, 12(3):889, 22 Jan 2023
Cited by: 2 articles | PMID: 36769537 | PMCID: PMC9917924
A 92 protein inflammation panel performed on sonicate fluid differentiates periprosthetic joint infection from non-infectious causes of arthroplasty failure.
Sci Rep, 12(1):16135, 27 Sep 2022
Cited by: 5 articles | PMID: 36167782 | PMCID: PMC9514711
Similar Articles
To arrive at the top five similar articles we use a word-weighted algorithm to compare words from the Title and Abstract of each citation.
The use of dual mobility cups in revision total hip arthroplasty for failed large head metal-on-metal bearings.
Int Orthop, 48(3):719-727, 01 Nov 2023
Cited by: 0 articles | PMID: 37907694 | PMCID: PMC10901945
In Revision THA, Is the Re-revision Risk for Dislocation and Aseptic Causes Greater in Dual-mobility Constructs or Large Femoral Head Bearings? A Study from the Australian Orthopaedic Association National Joint Replacement Registry.
Clin Orthop Relat Res, 480(6):1091-1101, 03 Jan 2022
Cited by: 2 articles | PMID: 34978538 | PMCID: PMC9263451
Dual Mobility Acetabular Cups in Primary Total Hip Arthroplasty in Patients at High Risk for Dislocation.
Surg Technol Int, 30:251-258, 01 Jul 2017
Cited by: 19 articles | PMID: 28395391
Outcomes of dual mobility components in total hip arthroplasty: a systematic review of the literature.
Bone Joint J, 100-B(1):11-19, 01 Jan 2018
Cited by: 98 articles | PMID: 29305445
Review