Abstract
Purpose
The low sensitivity and limitation to water phantoms of convection-dependent MRI magnitude signal-based proton beam visualization hinder its in vivo applicability in MR-integrated proton beam therapy. The purpose of the present study was, therefore, to assess possible contrast mechanisms for MRI phase signal-based proton beam visualization that can potentially be exploited to enhance the sensitivity of the method and extend its applicability to tissue materials.Methods
To assess whether proton beam-induced magnetic field perturbations, changes in material susceptibility or convection result in detectable changes in the MRI phase signal, water phantom characteristics, experiment timing, and imaging parameters were varied in combined irradiation and imaging experiments using a time-of-flight angiography pulse sequence on a prototype in-beam MRI scanner. Velocity encoding was used to further probe and quantify beam-induced convection.Results
MRI phase signal-based proton beam visualization proved feasible. The observed phase difference contrast was evoked by beam-induced buoyant convection with flow velocities in the mm/s range. Proton beam-induced magnetic field perturbations or changes in magnetic susceptibility did not influence the MRI phase signal. Velocity encoding was identified as a means to enhance the detection sensitivity.Conclusion
Because the MRI phase difference contrast observed during proton beam irradiation of water phantoms is caused by beam-induced convection, this method will unlikely be transferable to tightly compartmentalized tissue wherein flow effects are restricted. However, strong velocity encoded pulse sequences were identified as promising candidates for the future development of MRI-based methods for water phantom-based geometric quality assurance in MR-integrated proton beam therapy.Full text links
Read article at publisher's site: https://doi.org/10.1002/mrm.29752
Read article for free, from open access legal sources, via Unpaywall: https://onlinelibrary.wiley.com/doi/pdfdirect/10.1002/mrm.29752
Similar Articles
To arrive at the top five similar articles we use a word-weighted algorithm to compare words from the Title and Abstract of each citation.
MRI magnitude signal-based proton beam visualisation in water phantoms reflects composite effects of beam-induced buoyant convection and radiation chemistry.
Phys Med Biol, 68(18), 08 Sep 2023
Cited by: 0 articles | PMID: 37607554
Characterization of magnetic interference and image artefacts during simultaneous in-beam MR imaging and proton pencil beam scanning.
Phys Med Biol, 65(21):215014, 05 Nov 2020
Cited by: 4 articles | PMID: 33151908
Direct visualization of proton beam irradiation effects in liquids by MRI.
Proc Natl Acad Sci U S A, 120(23):e2301160120, 30 May 2023
Cited by: 0 articles | PMID: 37252953 | PMCID: PMC10265969
Integrating a low-field open MR scanner with a static proton research beam line: proof of concept.
Phys Med Biol, 63(23):23LT01, 22 Nov 2018
Cited by: 17 articles | PMID: 30465549