Abstract
Purpose
To evaluate the feasibility of CT angiography-derived fractional flow reserve (CT-FFR) calculations on ultrahigh-resolution (UHR) photon-counting detector (PCD)-CT series and to intra-individually compare the results with energy-integrating (EID)-CT measurements.Method
Prospective patients with calcified plaques detected on EID-CT between April 1st, 2023 and January 31st, 2024 were recruited for a UHR CCTA on PCD-CT within 30 days. PCD-CT was performed using the same or a lower CT dose index and an equivalent volume of contrast media. An on-site machine learning algorithm was used to obtain CT-FFR values on a per-vessel and per-patient basis. For all analyses, CT-FFR values ≤ 0.80 were deemed to be hemodynamically significant.Results
A total of 34 patients (age: 67.3 ± 6.6 years, 7 women [20.6 %]) were included. Excellent inter-scanner agreement was noted for CT-FFR values in the per-vessel (ICC: 0.93 [0.90-0.95]) and per-patient (ICC: 0.94 [0.88-0.97]) analysis. PCD-CT-derived CT-FFR values proved to be higher compared to EID-CT values on both vessel (0.58 ± 0.23 vs. 0.55 ± 0.23, p < 0.001) and patient levels (0.73 ± 0.23 vs. 0.70 ± 0.22, p < 0.001). Two patients (5.9 %) with hemodynamically significant lesions on EID-CT were reclassified as non-significant on PCD-CT. All remaining participants were classified into the same category with both scanner systems.Conclusions
While UHR CT-FFR values demonstrate excellent agreement with EID-CT measurements, PCD-CT produces higher CT-FFR values that could contribute to a reclassification of hemodynamic significance.Similar Articles
To arrive at the top five similar articles we use a word-weighted algorithm to compare words from the Title and Abstract of each citation.
Intra-individual comparison of coronary CT angiography-based FFR between energy-integrating and photon-counting detector CT systems.
Int J Cardiol, 399:131684, 25 Dec 2023
Cited by: 1 article | PMID: 38151162
Intraindividual Comparison of Ultrahigh-Spatial-Resolution Photon-Counting Detector CT and Energy-Integrating Detector CT for Coronary Stenosis Measurement.
Circ Cardiovasc Imaging, 17(10):e017112, 27 Sep 2024
Cited by: 0 articles | PMID: 39328060
Intra-individual comparison of epicardial adipose tissue characteristics on coronary CT angiography between photon-counting detector and energy-integrating detector CT systems.
Eur J Radiol, 181:111728, 07 Sep 2024
Cited by: 0 articles | PMID: 39276400
High-Resolution Head CTA: A Prospective Patient Study Comparing Image Quality of Photon-Counting Detector CT and Energy-Integrating Detector CT.
AJNR Am J Neuroradiol, 45(10):1441-1449, 03 Oct 2024
Cited by: 0 articles | PMID: 39237360
Funding
Funders who supported this work.