Continuous monitoring of spine movement function could enhance our understanding of low back pain development. Wearable technologies have gained popularity as promising alternative to laboratory systems in allowing ambulatory movement analysis. This paper aims to review the state of art of current use of wearable technology to assess spine kinematics and kinetics. Four electronic databases and reference lists of relevant articles were searched to find studies employing wearable technologies to assess the spine in adults performing dynamic movements. Two reviewers independently identified relevant papers. Customised data extraction and quality appraisal form were developed to extrapolate key details and identify risk of biases of each study. Twenty-two articles were retrieved that met the inclusion criteria: 12 were deemed of medium quality (score 33.4-66.7%), and 10 of high quality (score >66.8%). The majority of articles (19/22) reported validation type studies. Only 6 reported data collection in real-life environments. Multiple sensors type were used: electrogoniometers (3/22), strain gauges based sensors (3/22), textile piezoresistive sensor (1/22) and accelerometers often used with gyroscopes and magnetometers (15/22). Two sensors units were mainly used and placing was commonly reported on the spine lumbar and sacral regions. The sensors were often wired to data transmitter/logger resulting in cumbersome systems. Outcomes were mostly reported relative to the lumbar segment and in the sagittal plane, including angles, range of motion, angular velocity, joint moments and forces. This review demonstrates the applicability of wearable technology to assess the spine, although this technique is still at an early stage of development.

Free full text 


J Biomech. 2017 Nov 7; 64: 186–197.
PMCID: PMC5700811
PMID: 29102267

Wearable technology for spine movement assessment: A systematic review

Abstract

Continuous monitoring of spine movement function could enhance our understanding of low back pain development. Wearable technologies have gained popularity as promising alternative to laboratory systems in allowing ambulatory movement analysis. This paper aims to review the state of art of current use of wearable technology to assess spine kinematics and kinetics.

Four electronic databases and reference lists of relevant articles were searched to find studies employing wearable technologies to assess the spine in adults performing dynamic movements. Two reviewers independently identified relevant papers. Customised data extraction and quality appraisal form were developed to extrapolate key details and identify risk of biases of each study. Twenty-two articles were retrieved that met the inclusion criteria: 12 were deemed of medium quality (score 33.4–66.7%), and 10 of high quality (score >66.8%). The majority of articles (19/22) reported validation type studies. Only 6 reported data collection in real-life environments. Multiple sensors type were used: electrogoniometers (3/22), strain gauges based sensors (3/22), textile piezoresistive sensor (1/22) and accelerometers often used with gyroscopes and magnetometers (15/22). Two sensors units were mainly used and placing was commonly reported on the spine lumbar and sacral regions. The sensors were often wired to data transmitter/logger resulting in cumbersome systems. Outcomes were mostly reported relative to the lumbar segment and in the sagittal plane, including angles, range of motion, angular velocity, joint moments and forces.

This review demonstrates the applicability of wearable technology to assess the spine, although this technique is still at an early stage of development.

Keywords: Spine, Wearable sensor, Motion analysis, Kinematics, Kinetics

1. Introduction

Low back pain (LBP) has been identified as a major cause of adult disability contributing to substantial limitations in daily, recreational and occupational activities (Hoy et al., 2014). Despite LBP being a common problem, there is still a lack of understanding of what causes it in the absence of identifiable underlying pathology, making its management controversial. This is the case for non-specific LBP which accounts for up to 85% of all cases (Dankaerts et al., 2006). However, it is believed that the cause of LBP is multi-factorial involving both mechanical and psychosocial factors (Van Dillen et al., 2007, Clays et al., 2007). Whilst the latter are measured through questionnaires and scores reflecting patients’ subjective views, there is still an open debate on mechanical factors and their contribution. To obtain a better insight into the mechanical factors, an understanding of spine movement patterns during daily tasks has been recommended (Hernandez et al., 2017, Mitchell et al., 2008, Van Dillen et al., 2007). Daily tasks may impart, when impaired, abnormal, repetitive and prolonged stresses on the spine which may lead to LBP (Hernandez et al., 2017). Continuous monitoring of spinal movement provides the opportunity for objective and quantitative assssemnt of kinematics and/or kinetics which provides an insight into how spine movements influence mechanical changes and hence LBP development and persistence. In this way the link between LBP and daily activities can be assessed. However, recent recommendations advocate to not limit the analysis to the affected spine regions but to consider the whole body kinematic chain (McGregor and Hukins, 2009, Muller et al., 2015). Since lower limb kinematic/kinetic assessment with both laboratory and portable technologies is widely documented in the literature, this study is focused on spinal movement assessment.

Spinal movement analysis techniques have evolved mainly around the use of 3D motion tracking systems and electromagnetic tracking devices in a laboratory environment (Mitchell et al., 2008, Muller et al., 2015). These systems provide complex descriptions of body segment movement but only provide snap-shots of the subject over a short period of time, have limited capture volume constrained by camera positioning or receiver-magnetic field source distance, and create artificial environments for movement assessment. Therefore, despite the detailed information obtained these systems fail to reflect real-life situations.

Conversely, in-field measurements could aid assessment, objectify treatment pathways, and mitigate risks exposure. They will also support monitoring and treatment delivery thus facilitating and encouraging patients’ self-management. LBP is often related with the workplace and therefore is best understood through in-field measurements. These will facilitate our understanding of works ergonomics and the relationship between environment and LBP.

Advancements in miniaturised technology have brought about new possibilities for long-term monitoring of body movement including assessments in real-life environments. Portable and wearable sensors, in particular inertial sensors, have been introduced and have rapidly gained popularity in biomechanical studies of movement (Bonato, 2010). Properties such as light-weight, small size, low cost, energy efficiency and portability make these sensors suitable for a variety of applications, from simply monitoring activities of daily living (Yang and Hsu, 2010) or walking speed (Yang and Li, 2012) to more complex body segment kinematics estimation, particularly for the lower limbs (Fong and Chan, 2010), gait analysis (Shull et al., 2014) and balance assessment (Hubble et al., 2015).

In the context of LBP, and in the interest of capturing spine movement in real-life settings, this review aims to understand the state of art of current use of wearable technologies, for the assessment of spine kinematics and kinetics.

2. Methods

2.1. Search strategy

A systematic search of the following electronic databases was performed from inception up until August 2016: PubMed, Embase, ACM, Scopus and IEEEXplore. The search focused on identifying articles that included terms under the following general categories: sensors, wearable, outcome, spine. The search terms used are shown in Table 1. Hand searching and screening the reference lists of relevant articles were also performed to identify articles that may have been overlooked by the electronic searches. Retrieved articles were imported into EndNote X7 software (Thomson, Reuters, Carlsbad, CA).

Table 1

Search terms used in the systematic review.

GeneralSpecific search terms
SensorsSensor OR sensors OR sensing OR inertia OR inertial OR accelerometer OR gyroscope OR goniometer OR goniometry OR electrogoniometer OR “smart textile” OR “body sensor network”
AND
WearableWearable OR portable OR movable OR worn OR ambulatory OR “non-invasive” OR “body-mounted”
AND
OutcomeKinetic OR kinetics OR kinematic OR motion OR motions OR movement OR assessment OR “joint angle”
AND
SpineSpine OR spinal OR back OR cervical OR thoracic OR lumbar OR vertebra

2.2. Inclusion and exclusion criteria

The titles and abstracts of articles retrieved from the searches were assessed independently by two reviewers (EP, WSK). Full text of potential articles was assessed against eligibility criteria by the two reviewers independently. Inclusion and exclusion disagreements were resolved by consensus. Articles were included if they satisfied the following criteria: were published in English; assessed the spine, including the lumbar spine, using wearable technologies; were peer-reviewed; included at least one of following as outcome measures: spine kinematics, kinetics, and posture parameters as obtained from wearable technology; involved an adult population (≥18 years old) performing dynamic movement tasks. Articles were excluded if: they were review or case-study; they used only non-wearable devices; wearable technology was used to only quantify physical activity or spatio/temporal parameters of the activity performed; they described a potential technology not validated/used with human subjects; they did not assess motion of the spine.

2.3. Data extraction and quality appraisal

Customised data extraction and quality appraisal forms were developed to extrapolate key details and identify risk of biases of each study. The following details were extracted: study design, sample size, participants’ demographics (e.g. population type, age, gender, mass, height), tasks conducted, measuring system used, data sampling, participant set-up (e.g. positioning of the sensors, fixation method), data processing (e.g. filter use for the signal), kinematic and kinetic variables evaluated from sensors signals, statistical analysis technique, reliability/accuracy evaluation.

The quality of the selected studies was assessed using a customised checklist as no applicable standardised guidelines were identified. The checklist was however developed based on tools previously used in motion analysis reviews to include elements related to external validity (e.g. sampling methods and participants description), and biases in protocol description and outcomes reporting, with specific questions to assess information relating to technology use and signal evaluation (Dobson et al., 2007, Needham et al., 2016). The quality checklist consisted of 20 items (Table 2); each item was rated as zero (no detail), one (limited detail) and two (good detail).

Table 2

Quality assessment results of included articles.

Quality Index Item numberBartalesi et al. (2010)Baten et al. (1996)Boocock et al. (1994)Charry et al. (2011)Chhikara et al. (2010)Consmuller et al. (2012)Dunk and Callaghan (2010)Faber et al. (2016)Goodvin et al. (2006)Ha et al. (2013)Khurelbaatar et al. (2015)Lee and Park (2011)Lee et al. (2011a)
12222222222222
20010100002220
30010022202222
40021022022012
50000010000000
60000000000000
72122222222122
80010000002000
90021122221021
101122221122122
112222222122122
122221220122122
132121222122222
142112112111122
151222212122222
162222200222200
172222210212211
181122222222222
181022222212222
200000120202211


Total score (/40)20173024262823242534252927
Percentage score50437560657058606385637368
Quality categoryMMHMMHMMMHMHM
Quality Index Item numberLee et al. (2011b)Morlock et al. (2000)Nevins et al. (2002)O’Sullivan et al. (2012)Thoumie et al. (1998)Van Hoof et al. (2012)Walgaard et al. (2016)Wong and Wong (2008a)Wong and Wong (2008b)Paper with full scorePaper with half scoreTotalHalf score paper %Full score paper %Total %
1222222222220220100100
20202111115712322355
32202222221511656873
42102121009514234164
501010200013414518
6020000000101055
7222212222193221486100
80200120002249918
91022012009615274068
10221222122157223268100
11221222222193221486100
1222222222218321148296
1322021212216521237396
14221111211814226436100
15221222222184221882100
161202222221611757377
1711122221213821365995
18221222222193221486100
182202222221822098291
201202022229413184059


Total score (/40)283514342635302728
Percentage score708835856588756870
Quality categoryHHMHMHHMH

M:Medium, H:High.

1. Were the research objectives or aims clearly stated?

2. Was the study design clearly described?

3. Was the study population adequately described?

4. Were the eligibility criteria specified?

5. Was the sampling methodology appropriately described?

6. Was the sample size used justified?

7. Did the method description enable accurate replication of the measurement procedures?

8. Was the participants’ assessor described (e.g. expertise)?

9. Was a system for standardizing movement instructions reported?

10. Was the equipment design and set up clearly described?

11. Were sensors locations accurately and clearly described?

12. Was sensor attachment method clearly described?

13. Were the spine segments analysed clearly described?

14. Was the signal/data handling described?

15. Were the main outcomes measured and the related calculations (if applicable) clearly described?

16. Was the system compared to an acknowledged gold standard?

17. Were measures of reliability/accuracy of the equipment used reported?

18. Were the main findings of the study stated?

19. Were the statistical tests appropriate?

20. Were limitations of the study clearly described?

3. Results

3.1. Articles selection

The search identified 1837 potentially relevant articles with 15 articles identified from references of related articles and hand searches. 1610 articles remained for consideration after removing duplicates. Following screening of title and abstract, full texts of 46 articles were retrieved. Twenty-two articles satisfied the inclusion criteria. The articles selection process and reasons for full-text articles exclusion are shown in Fig. 1.

An external file that holds a picture, illustration, etc.
Object name is gr1.jpg

Flow diagram depicting the review process.

3.2. Quality of reviewed articles

The overall quality of papers was rated according to Ratcliffe et al. (2014), 12 papers were deemed of medium quality (score 33.4–66.7%), and 10 of high quality (score >66.8%) (Table 2). There was a significant low rate of addressing a few core methodological questions: sample size was justified in only 1 paper; 4 papers described sampling methodology; assessor expertise was reported in 4 papers; participants’ eligibility criteria were fully described in 9 and partly in 5 papers, standardised movement instructions and limitations were reported in 14 and 13 papers respectively. Higher quality, on the other hand, was generally recorded in reporting research objectives, and in the description of the methods and systems used. Comparison with a gold standard system was described in 17 papers and measures of accuracy/reliability of the system used in 21 papers.

3.3. Descriptive aspects of reviewed articles

In three studies (Dunk and Callaghan, 2010, Lee et al., 2011b, Van Hoof et al., 2012), the aim was to compare spinal motion differences between healthy and LBP groups. All other studies aimed at introducing and/or validating portable assessment tools for spine movement evaluation. The sample size of the papers reviewed was small with 10 of them considering up to 10 participants (Baten et al., 1996, Bartalesi et al., 2010, Charry et al., 2011, Chhikara et al., 2010, Faber et al., 2016, Goodvin et al., 2006, Khurelbaatar et al., 2015, Lee and Park, 2011, Nevins et al., 2002, Wong and Wong, 2008a), including 4 in which only one participant was involved. In only 4 studies, the sample size was greater than 25 (Consmüller et al., 2012, Dunk and Callaghan, 2010, Ha et al., 2013, Walgaard et al., 2016).

Most of the studies assessed spine motion in healthy young adult subjects. Five studies also recruited LBP participants (Dunk and Callaghan, 2010, Lee et al., 2011a, Lee et al., 2011b, Morlock et al., 2000, Van Hoof et al., 2012).

Range of motion (ROM) manoeuvres were common tasks performed by the participants, however tasks such as lifting (2/22) (Baten et al., 1996, Charry et al., 2011), seated computer use (1/22) (Dunk and Callaghan, 2010), cycling (1/22) (Van Hoof et al., 2012), walking (2/22) (Khurelbaatar et al., 2015, Lee and Park, 2011), sit-to-stand (2/22) (Walgaard et al., 2016, Wong and Wong, 2008b) and stairs climbing (2/22) (Lee and Park, 2011, Lee et al., 2011b) were also considered. Some studies monitored participants in their working environments (3/22) (Boocock et al., 1994, Morlock et al., 2000, Thoumie et al., 1998) and during everyday activities in real-life contexts (3/22) (Nevins et al., 2002, Wong and Wong, 2008a, Wong and Wong, 2008b) and simulated in laboratories (3/22) (Goodvin et al., 2006, Nevins et al., 2002, O’Sullivan et al., 2012).

3.4. Sensors specification and set-up

Details of sensors described in each paper are reported in Table 3.

Table 3

Sensors specifications, set-up and outcomes recorded for the included articles.

ArticlesUnitsTypeSize and WeightData Transmission/StorageSample FrequencyData FilteringSensors PlacementSpine Outcomes reportedGold Standard
Bartalesi et al., 20103Newly developed system composed by: textile piezoresistive sensor (single strip of conductive elastometer) and 2 tri-axial accelerometers (LIS3L02Al by ST Microelectronics)128 th order Finite Impulse Response (FIR) filterAccelerometers: T12, Sacrum.
Piezoresistive strip over the lumbar spinous processes
Length of lumbar archBTS Elite4 optical motion capture system (BTS Bioengineering Corp., New York, USA)
Baten et al.,19961Accelerometer (IC sensors™), miniature rate gyroscope (Murata™), sensors mounted on solid strip of orthoplast materialL5/S1Actual back absolute angle; absolute inclination of the line between L5/S1 and the shoulder from gyroscope signal; semi-stationary estimation of the absolute inclination of the same line as aboveVicon optical motion capture system (Oxford Metrics, Oxford, UK)
Boocock et al., 19942Flexible electrogoniometer (2 planes measurement) (Penny and Giles, Blackwood, UK)48 gData logger: 156 g pocket sized module; 65 K memory (27.3 min continuous data storage)20 HzOne end on spinous process of S1 and S2, and second end over T12Sagittal angle of lumbar curvatureFluid-filled inclinometer (MIE Medical Research Ltd. Leeds, UK), and draftsman's flexicurve (Burton 1986)
Charry et al., 20112Newly developed system ‘ViMove’ composed by: 3D accelerometer (Bosch SMB380) with Serial Peripheral Interface (SPI) digital output and low noise 1D gyroscope (Epson XV8100)Wireless to base station (256 Mb)20 Hz2nd order Butterworth filter, 5 Hz cut-off frequencyL1, S13D lumbar anglesNDI Optotrack System (Northen Digital, Waterloo ON, Canada)
Chhikara et al., 20102Newly developed system ‘IMPAIRED’ composed by: [BSN Development Kits v3] tri-axial accelerometer (Analog Devices ADXL330), tri-axial gyroscope formed by 2 perpendicularly mounted dual axis gyroscopes (Invensense IDG300A) only one axis connectedWireless to Base station, 15 m range, 1 h battery life25 HzL1, L5/S1Lumbar and pelvic angular velocities, Lumbar and pelvic angles and range of motions (sagittal plane of movement)NDI PolarisVicra optical tracking system
(Northen Digital, Waterloo ON, Canada)
Consmuller et al., 20122Epionics SPINE (Epionics Medical GmbH, Potsdam, Germany): Strain-gauge sensor strips with 3D accelerometer at lower end of each strip120 gData logger wired to sensors50 HzParavertebrally 5 cm away from spinal column with caudal segment on PSISLordosis/kyphosis angle, Lumbar range of motion (sagittal plane of movement)
Dunk and Callaghan, 20102Tri-axial accelerometer (S2-10G-MF, NexGen Ergonomics, Montreal, Quebec, Canada)128 Hz4 th order Butterworth filter, 1 Hz cut-off frequencyL1, S2Lumbar spine angle, Lumbar range of motion, Shift (step-like angular change in lumbar angle), Fidgets(small rapid changes of the lumbar angle about the same average position)
Faber et al., 201617Xsens MVN full body inertia/magnetic motion capture system (Xsens technologies BV, Enschede, Netherlands)120 HzPelvis, Head, upper arms, forearm, thighs, shank, feet, scapulae, sternum, hand3D L5/S1 moment time series and peaks; ground reaction forcesCertus Optotrack optical motion capture system (Northen Digital, Waterloo ON, Canada)
2 Kistler force platforms (Kistler Instrumente AG, Winterthur, Switzerland)
Goodvin et al., 20063MT9 sensors (Xsens technologies BV, Enschede, Netherlands)XBus device wired to sensors, data transferred via serial cable to PC from XBusHead, C7/T1, L4-5Cervical, thoracic and lumbar flexion/extension and lateral bending angles, roll, pitch and yaw motionsVicon optical motion capture system (Oxford Metrics, Oxford, UK)
Ha et al., 20132MTx sensors (Xsens technologies BV, Enschede, Netherlands)38 × 53 × 21 mm; 30 gXBus wired to sensors, Bluetooth data transfer to a PC from XBus100 HzL1, S1Lumbar 3D range of motionElectromagneticFastrak system (Polhemus, USA)
Khurelbaatar et al., 201517Xsens MVN full body inertia/magnetic motion capture system (Xsens technologies BV, Enschede, Netherlands)
Pedar-X in-shoe pressure measurement system (Novel GmbH, Munich, Germany)
Head, trunk, pelvis, upper and lower extremitiesCervical, thoracic, lumbar joint forces, joint momentsHawk® motion capture system (Motion Analysis, Santa Rosa, CA, USA)
MP4060® force platform (Bertec Corporation, Columbus, OH, USA)
Lee and Park, 20115MTx sensors (Xsens technologies BV, Enschede, Netherlands)XBus wired to sensors, Bluetooth data transfer to a PC from XBus100 HzXsens Kalman filter algorithmT1, T12, S1, and 2 sensors 6 cm above each ankle in alignment with the fibulaThoracic, Lumbar, Pelvis 3D angles, range of motions
Lee et al., 2011a3MTx sensors (Xsens technologies BV, Enschede, Netherlands)XBus wired to sensors, Bluetooth data transfer to a PC from XBus50 HzXsens Kalman filter algorithmT1, T12, S1Thoracic, Lumbar, Pelvis flexion/extension angles, range of motions, angular velocities
Lee et al., 2011b53 MTx sensors and 2 MT9 (Xsens technologies BV, Enschede, Netherlands)XBus wired to sensors, Bluetooth data transfer to a PC from XBus50 Hz
100 Hz
Xsens Kalman filter algorithmT1, T12, S1 and two over
Shank above each ankle
Thoracic, Lumbar, Pelvis 3D angles, range of motion
Morlock et al., 200081 Uni and 7 bi-axial goniometers (Penny and Giles, Gwent, UK)
Pedar in-shoe pressure measurement system (Novel GmbH, Munich, Germany)
24-channel data logger (BIOSTORE, Wehrheim, Germany) connected via cables to the goniometers, RAM card with 45 min max duration.
Pedar data stored onto the hard-disk of a laptop connected to the subject via a 10 m cable.
10 Hz
20 Hz (Pedar system)
4 th order Butterworth filter, 1.8 Hz cut-off frequencyInferior of S1, Superior of T12,
Lower limbs
3D lumbar spine angles, 3D joint moment at L5/S1, Bone to Bone contact force at L5/S1Vicon optical motion capture system (Oxford Metrics, Oxford, UK)
O’Sullivan et al., 20121BodyGuard™ spinal posture monitoring device (Sels Instruments, Belgium)
(Strain gauge unit)
Battery powered processing unit, wireless communication to PC20 HzAcross L3 and S2Lower lumbar extension/flexion as % strain gauge elongation, posture as % range of motion: mean and peak valuesCartesian Optoelectronic Dynamic Anthropometer (CODA™) mpx64 (Charnwood Dynamics Ltd, Leicestershire UK)
Thoumie et al., 199822D Electrogoniometer (M180, standard version, Penny and Giles, Gwent, UK)2 × 5 cm
2 × 14 cm
(End plates measures)
Digital recorder (Angle display unit, Penny and Gilles, Gwent, UK) orPortable data logger (DL 1001, Penny and Gilles, Gwent, UK)
with max 163 min of recording at 3 Hz
3 HzOne end plate over S1 and the second plate 10 cm apart over the lumbar spineLumbar flexion/extension angle maximum value and range of motion, Mean lumbar lordosis curvatureX-ray
Van Hoof et al., 20122BodyGuard™ spinal posture monitoring device (Sels Instruments, Belgium)
(Strain gauge unit)
10 × 28 mm measure of end pieces of strain gauge unitSignal processing unit (56 × 71 × 15 mm). Data storage max 24 h20 HzL3, S2Lower lumbar flexion/extension angle as % of total lumbo-pelvic range of motionCartesian Optoelectronic Dynamic Anthropometer (CODA™) (Charnwood Dynamics Ltd, Leicestershire UK)
Walgaard et al., 20161Single inertia sensor comprised of 3 accelerometers and 3 gyroscopes (DynaPort® Hybrid, McRoberts)100 Hz2nd order Butterworth filter, 15 Hz cut-off frequencyL4Lumbar flexion range of motion, maximum flexion velocity, maximum forward velocity, forward velocity during seat-off and heel strike, maximum vertical velocity, vertical velocity at heel strike, 3D accelerations, velocities, displacements and anglesCertus Optotrack optical motion capture system (Northen Digital, Waterloo ON, Canada)
Wong and Wong, 2008a3Newly developed system, each sensor module composed by: Tri-axial acceleromter (KXM52-Tri-axis, Kionix) and 3 uni-axial gyroscopes (Epson)Sensor module: 22 × 20 × 12 mm each 6 g
Battery holder: 50 × 55 × 12 mm, 82 g, 8 h recording
Digital data acquisition and feedback (buzzer) system (21 × 50 × 84 mm, 44.5 g)T1/2, T12, S1 (Elastic garment)Average angles of thoracic and lumbar curves in the sagittal and coronal planesVicon optical motion capture system (Oxford Metrics, Oxford, UK)
Wong and Wong, 2008b3Newly developed system:each sensor module composed by: Tri-axial acceleromter (KXM52-Tri-axis, Kionix) and 3 uni-axial gyroscopes (Epson)Sensor module: 22 × 20 × 12 mm each 6 g
Battery holder: 50 × 55 × 12 mm, 82 g, 8 h recording)
Digital data acquisition and feedback (buzzer) system (21 × 50 × 84 mm, 44.5 g)5th order Butterworth filter, 4 Hz cut-off frequencyT1/2, T12, S1 (Elastic straps)Thoracic and lumbar angular velocities, thoracic and lumbar peak angles in the sagittal and coronal planesVicon optical motion capture system (Oxford Metrics, Oxford, UK)
Nevins et al., 20026Newly developed system: sensors: Analog Devices model ADXL202E comprising 2 accelerometers in perpendicular axes and 6 satellite processors (AT90S2313)Sensors and harness 125 gData logger, 32 MB compact flash card, 69 × 115 × 38 mm; 250 g15 s epochsPositioned along the vertical axis of the spine, exact location not specifiedAngles relative to the vertical in each sensor in the sagittal plane

Newly developed and off-the shelf wearable sensors were used within the reviewed studies.

The type of sensors used to assess spine movement varied from electrogoniometers (3/22), strain gauges based sensors (3/22), textile piezoresistive sensor (1/22) and uniaxial to triaxial accelerometers, the latter often used in conjunction with gyroscopes and magnetometers (15/22). Two studies also used portable instrumented insoles in addition to the above motion sensors.

The number of sensor units used differed across the papers, with on average two sensor units positioned on the back of the participants. Positioning along the spine varied with the majority of the included studies reporting sensors placement over lumbar and sacral landmarks (L1, L3, L4, L5, S1, S2) (18/22) and T12 (8/22). Nine studies also reported positioning of the sensors on different regions of the back or body parts.

Different fixation methods were adopted, including single and double-sided adhesive tape, velcro strap, elastic belt, and neoprene strap. Some sensors were also attached using a rigid support or embedded into clothing. The size and weight of the systems used was reported in only few studies. Authors reported sampling frequencies between 3 Hz and 100 Hz, however not all authors specified the frequency used.

3.5. Data storage and processing

Data were collected and stored for subsequent processing in the majority of studies. Only one study described the use of real-time feedback (Goodvin et al., 2006). Data from the sensors were wirelessly transmitted to PCs/base station for data storage by the sensor units directly or, by data loggers wired to the sensors, which communicate via Bluetooth to the storage station. In eight studies, data were saved in laptops/portable data loggers attached on the participants’ bodies and wired to the sensors units (Boocock et al., 1994, Consmüller et al., 2012, Morlock et al., 2000, Nevins et al., 2002, Thoumie et al., 1998, Van Hoof et al., 2012, Wong and Wong, 2008a, Wong and Wong, 2008b). The capacity of the memory cards of the portable data loggers was seldom described.

Most of the systems required calibration before use and filtering due to noise in the data collected. This adds time to the data processing procedure before calculation of the outputs. Butterworth filters of the 2nd to the 5 th orders were adopted with cut-off frequencies ranging from 1 Hz to 30 Hz (6/22, Table 3). Kalman filters were also used (3/22) and one study used a 128 th order finite impulse response filter (Table 3). The rest of the studies did not report the type of filter used.

The sensors signals were processed to calculate different kinematic and kinetic aspects of spine movements in 3D. More commonly, however, only sagittal plane movement data were reported. Outcomes were mostly reported relative to the lumbar segment, although cervical, thoracic spine segments, and pelvis were considered in some studies. The following outcomes were reported: 3D angles, flexion/extension ROM, spine curvature sagittal angles, angular velocity, joint moments, and joint forces. Details of outcome measures for each study are shown in Table 3.

3.6. Reliability and accuracy

The accuracy and reliability of the systems proposed in the reviewed papers were evaluated by comparisons with gold standard systems. Optoelectronic motion capture systems were mostly used as gold standards (15/22), one study used an inclinometer (Boocock et al., 1994), and one used X-Ray (Thoumie et al., 1998). Direct comparison with a gold standard was not performed in six studies (Consmüller et al., 2012, Dunk and Callaghan, 2010, Lee and Park, 2011, Lee et al., 2011a, Lee et al., 2011b, Nevins et al., 2002). Comparisons among studies in terms of reliability and accuracy of systems used are difficult to assess as different measures were reported, they were calculated on different outcome measures and during different test circumstances. Measures of reliability and accuracy reported are shown in Table 4 for each paper.

Table 4

Accuracy and reliability of systems described in the reviewed articles.

ArticlesAccuracy/Reliability
Bartalesi et al., 20102% error in length estimation; high correlation greater than 0.8 when comparing lumbar arch from reference system and new system
Baten et al.,1996Calibration error <1%; Over 8 h accelerometer offset drift ±5%, inclination depending error 3–20% in the semi-stationary estimation of the absolute inclination; error in absolute inclination from gyroscope ±10%
Boocock et al., 1994Calibration rig test results: 0.96 between systems angles; RMSE of 2.5° equivalent to 6° limit of agreement; Crosstalk error between 7% and 10% of the measured angle. Electrogoniometer 5.7° RMS difference, 0.78 correlation, 1.17° mean difference, intrasubject SD 4.05° between two different test occasions; Recording angle error <3 compared to gold standards; Electrogoniometer vs Fluid-filled inclinometer: RMS difference 3.89°, correlation 0.9, mean difference 1°; Electrogoniometer vs draftsman's flexicurve: RMS difference 5.87°, correlation 0.77, mean difference −1.17°
Charry et al., 2011The RMSE achieved for one dimensional movements in the Flexion, Lateral Flexion and Twist planes were 1.0°, 0.5° and 2.4° respectively, 2.0°, 3.1°, and 5.1° for 2D movements and 2.1°, 2.4° and 4.6° for 3D movements. RMS errors averaged over the 53 movements performed by two test subjects: 1.9° and 2.1° for Flexion 2.4° and 2.1° for Lateral Flexion and 5.2° and 4.1° for Twist for Subject 1 and 2 respectively
Chhikara et al., 2010Mean average error for angular velocity: (Lumbar) 1.52 ± 31.24, (Pelvis) 0.78 ± 9.4. Mean average error for angles: (Lumbar) −1.83 ± 1.85, (Pelvis) 0.91 ± 0.28; error in orientation results between system between 3 and 7°
Consmuller et al., 2012Good to excellent correlation between the left and right sensors for all angles in the sagittal plane with average Pearson correlation coefficient r = 0.81. The correlation for upright standing was r = 0.85, r = 0.70 for maximum flexion, and r = 0.87 for maximum extension. The correlation for the repeated measurements on three different days was very good for segmental results during upright standing (ICC = 0.87), flexion (ICC = 0.86) and extension (ICC = 0.84), with similar results for lordosis (ICC = 0.85), flexion (ICC = 0.83) and extension (ICC = 0.79) angles. The average correlation coefficient was 0.84
Dunk and Callaghan, 2010No accuracy/reliability measures reported
Faber et al., 2016RMS errors were below 10 Nm for flexion, lateral flexion and twist L5/S1 moments time series. R^2 values > 0.993 for flexion L5/S1 moment and below 0.993 for lateral flexion and twist L5/S1 moments. ICC of the absolute peak moments were 0.971, 0.781 and 0.69 for flexion, lateral flexion and twist respectively.
3D GRF RMS errors remained below 20 N, R^2 above 0.981 for vertical GRF, around 0.6 for anterio/posterior and medio/lateral GRF. ICC of the vertical GRF peaks between systems was 0.998, and 0.948 and 0.559 for anterio/posterior and medio/lateral GRF peaks. During fast trunk bending overestimation of about 15 N in vertical GRF peak
Goodvin et al., 2006From head sensor: roll, pitch and yaw average deviation 0.1°, 0.42°, and 0.2° respectively. From torso sensor: roll, pitch and yaw average deviation 0.03°, 0.06° and 0.23° respectively. For hip sensor: roll, pitch and yaw average deviation 3.1° roll, 0.33° and 1.35° respectively
Ha et al., 2013Overall ROM R^2 0.999 between systems; flexion R^2 = 0.7878 and correlation coefficient = 0.8876; extension R^2 = 0.4321 and correlation coefficient = 0.6573; right lateral flexion R^2 = 0.7285 and correlation coefficient = 0.8535; left lateral flexion R^2 = 0.8101 and correlation coefficient = 0.900; right axial rotation R^2 = 0.4199 and correlation coefficient = 0.6657; left axial rotation R^2 = 0.2633 and correlation coefficient = 0.5411. Mean differences between −0.81 and −1.26°. No significant difference between systems p-values greater than 0.05
Khurelbaatar et al., 2015Bench test: single axis rotation: average RMSE 0.9 ± 0.7°; 3-axis rotation: average RMSEs were 0.8 ± 0.68° in the X-axis, 1.1 ± 0.58° in the Y-axis, and 0.8 ± 0.58° in the Z-axis.
GRF normalised RMSE < 9.1% and Pearson's correlation r ≥ 0.96; Lumbar joint force normalised RMSE 5.8% and r = 0.81; Cervical joint force normalised RMSE 6.0%, r = 0.81; Thoracic joint force normalised RMSE 6.0% and r = 0.80. Moments Lumbar joint: normalised RMSE 16.9% and r = 0.86; Cervical joint: normalised RMSE 13.6% and r = 0.8; Thoracic joint normalised RMSE 8.4% and r = 0.93
Lee and Park, 2011Not measured directly in current study. 2° RMS in dynamic motion from sensors manufacturer; Orientation error <3.1° from previous study from the same authors using MT9 sensors (which are an older version of the MTx used in this study) against a Vicon optical motion capture system
Lee et al., 2011aNot measured directly in current study. 2° RMS in dynamic motion from sensors manufacturer
Lee et al., 2011bNot measured directly in current study. 2° RMS in dynamic motion from sensors manufacturer. Orientation error 3.1° in roll, 0.3° in pitch, and 1.4° in yaw for MT9 sensors as measured in previous authors' study against Vicon optical motion capture system
Morlock et al., 2000Proposed system overestimates max value of compressive Bone-to-bone (B-t-B) force by 9 ± 4%; the medio-lateral BtB force by 3 ± 58%; the antero-posterior BtB 23 ± 4%; flexion/extension moment by 12 ± 4%; the lateral flexion moment by 36 ± 7%, and the torsional moment by 67 ± 14%. Lumbar spine angles flexion/extension 1.8 ± 1% difference between systems; for torsion and lateral flexion plane the proposed system overestimated angles by more than 100%
O’Sullivan et al., 2012Strong positive correlations and small differences between systems in sitting (Spearmans rank correlation coefficient rs = 0.88 and R^2 = 0.78; difference 2.39°) and standing (rs = 0.88 and R^2 = 0.78; difference 3.06°). Overall mean difference in standing and sitting <10%ROM. Agreement varied among a range of tasks in sitting and standing with differences up to a maximum of 6.2° in sit-to-stand flexion and 5.8° when putting shoe on
Nevins et al., 2002Precision ±0.39° during static condition
Thoumie et al., 1998Electrogoniometers repeatability ±2° between +90 and −90°; Cross talk between 2 electrogoniometers never exceeded 1° for lateral bending of less than 30°. Electrogoniometric and radiographic lumbar curve angles correlation in flexion, extension and standing positions were 0.76, 0.77 and 0.58 respectively but, the exact values were significantly different. Lumbar ROM correlation in flexion, extension and standing positions were 0.65, 0.74 and 0.48 respectively with values significantly different
Van Hoof et al., 2012Intra- and inter-rater reliability: ICC values: 0.837–0.874 and 0.914–0.940 respectively based on previous referenced study. rs = 0.88 and R^2 = 0.78, mean difference < 10% Flexion ROM according to another study. Additionally, the correlation between the two systems during ergometer cycling was evaluated in advance and was strong (r = 0.8), with a mean difference of 3°
Walgaard et al., 2016RMSE < 10% for 3D lumbar accelerations, velocities, displacements and angles between systems except for sideway displacement, and non-sagittal plane rotation with RMSE 40.1 ± 47.4%. ICC ≥ 0.867 for lumbar flexion range of motion, maximum flexion velocity, maximum forward velocity, forward velocity during seat-off and heel strike, maximum vertical velocity, except for vertical velocity at heel strike ICC = 0.649. Mean absolute differences were 0.45 ± 0.35° for flexion range, 16.9 ± 16.6° for maximum flexion velocity and 0.1 ± 0.06 m/s or lower for other velocity measures
Wong and Wong, 2008a, Wong and Wong, 2008b(Bench test) Static measurement: RMS difference <1°; Pearson’s correlation coefficient for angles >0.999. Dynamic tilting measurement: RMS difference <1.5°; Pearson’s correlation coefficient for angles >0.999, RMS angular velocity along the x-axis 35.2 ± 1.9°/s and along y-axis 34.1 ± 1.7°/s
Wong and Wong, 2008a, Wong and Wong, 2008b(Bench test) Static measurement: RMS difference <1°; Pearson’s correlation coefficient for angles >0.999. Dynamic tilting measurement: RMS difference <1.5°; Pearson’s correlation coefficient for angles >0.999, RMS angular velocity <40°/s.
(Participants tests) RMS differences between systems were <3.1° for the sagittal plane and <2.1° for the coronal plane at both thoracic and lumbar regions and movements (flexion, left and right lateral bending, stand-sit-stand). The correlation coefficients of the measurements were all above 0.829

4. Discussion

Use of wearable sensor based motion capture systems for the quantitative assessment of human movement is rapidly growing. Portable systems present as a practical choice for in-field measurements, thereby enabling analysis in real-life settings where movement disorders may develop and perpetuate. This is the case with LBP, notably one of the most frequent claimed occupational disorders (Kerr et al., 2001, Dagenais et al., 2008). In-field assessment, therefore, may positively impact on the healthcare provision of LBP through enhanced understanding of the underlying mechanics.

This systematic review aimed to examine the literature to understand the state of art in relation to the use of wearable technology to assess spine kinematics and kinetics, with the view to use them for clinical trials and work-based assessments. Twenty-two articles were found to satisfy the inclusion criteria. The appraisal of studies quality revealed that the overall scientific quality of the reviewed papers is of medium to high quality. However, it was found that papers were lacking a thorough description of their sample in terms of calculation and justification of the sample size, participants’ inclusion criteria, and assessors’ expertise. The use of standardised instructions for the experiments was unclear. On the other hand, studies performed well in elements related to the description of the technology used, experimental protocol and outcomes.

The small number of existing literature found on portable assessment of spine motion revealed limited adoption of this technique, which is at an early stage of development and translation. The majority of papers, reported validation type studies with a proof-of-concept theme. New and off-the-shelf technologies were described and validated against more traditional motion capture systems to assess spine movement, as a first step to demonstrate the potential of such technologies for clinical and daily use. Validation mainly occurred with small groups of healthy young participants; only 5 studies explored wearable sensors use with clinical populations. Moreover, the majority of studies were conducted in research laboratories, although some papers described data assessment in working environments and daily living. Boocock et al. (1994) used a flexible electrogoniometer to track changes in the lumbar spinal sagittal curvature in 4 garage mechanics over 2 h of their working day. An electrogoniometer was also used by Thoumie et al. (1998) to assess changes in lumbar flexion/extension when nurses and physiotherapists wore a belt during daily shifts. In Morlock et al. (2000) a full body portable system comprising of 8 electrogoniometers and pressure distribution insoles was used to determine lumbar spine loads and 3D lumbar angles in nurses, with and without LBP, during 4 h of continuous monitoring. A smart garment integrating a tri-axial accelerometer and 3 uni-axial gyroscopes for posture training was developed and used in two studies (Wong and Wong, 2008a, Wong and Wong, 2008b) with 5 healthy participants over 4 and 3-day trials during everyday activities. Sagittal and coronal plane angles of the thoracic and lumbar spine were estimated in these studies. Finally, Nevins et al. (2002) monitored spinal posture in one subject while day-to-day activities took place. These exceptions demonstrate, although still at a preliminary stage, the possibility to use these systems for in-field measurements.

When studies were performed in research laboratories, participants performed mostly ROM tasks even though more dynamic activities (e.g. walking) were assessed. The validity of using such technologies in dynamic everyday tasks is important for translation into real-life use. Such technologies in fact may reveal impractical during dynamic tasks in daily living in terms of positioning and sustainability of sensor attachment that will be critical for the robustness of the outputs. Moreover, the presence of wires may cause discomfort for long-term use and restrict natural movements, jeopardising data acquisition. Wires were used in the systems described to connect the sensor units to data loggers or directly to PCs compromising their wearability and practical translation. Alternatively, Bluetooth connection was used to transfer data to a stationary PC. In those cases, however, the systems cannot be considered entirely portable as a recording station is still required and the subject will always need to be within a certain distance to maintain Bluetooth connectivity. This could be obviated if a data logger is used. Data loggers, hence, represent comprises for portability despite adding substantial volume to the measurement system worn. Data logging is one area to be improved in the development of such technology in terms of miniaturisation, Bluetooth connectivity with the sensor units and memory capacity.

In the majority of studies, the systems used consisted of inertia-based sensors in the form of accelerometers, single or three axes, used alone or together with gyroscopes or gyroscopes and magnetometers in off-the-shelf inertia measurement units. Accelerometers and gyroscopes are known to be affected by drift errors when acceleration and angular velocity signals are integrated to obtain positions and orientations values, whilst magnetometers signals can be distorted by surrounding magnetic fields (Alonge et al., 2014). Other sources of errors are related to sensors positioning: skin movement artefact and misalignment between the sensor axes and underlying anatomical segments. The other types of systems described, electrogoniometer and strain gauges based sensors, are also sensitive to the correct sensor positioning, yet not affected by the presence of magnetic fields and post-processing analysis as they provide direct measures of angles. Despite this, technology developments have favoured the uptake of inertia sensors due to their small size and versatility. Moreover, spine movement is not isolated to one anatomical plane. Inertia sensors facilitate the characterisation of LBP related movement disorders by allowing 3D analysis. Despite the potential errors outlined, all systems showed good to excellent agreement and correlation with gold standards for the outcomes evaluated. Gold standards were represented, in the majority of the included studies, by optoelectronic motion capture systems which are frequently used to assess human movement. Measures of sensor accuracy and reliability reported included the root mean square error, mean difference, intraclass correlation coefficient, Pearson correlation coefficient, Spearman’s rank correlation coefficient and coefficient of determination. From the detailed values reported in Table 4, it can be recognised that all the systems could be used to measure kinematics and kinetics parameters of the spine with good accuracy and reliability. It should be mentioned that some of these measures were reported to be related to simple tasks or bench tests. It has, however, been shown that errors can be task dependent (O’Sullivan et al., 2012) therefore it is important to evaluate the accuracy in dynamic and extreme end of range movements prior to widespread implementation. Significant errors could arise from sensor malpositioning, poor fixation, and incorrect identification of anatomical landmarks. It is important to note that such errors also impact on optical tracking systems (Leardini et al., 2005). Currently, there are no robust tools to correct for movement artefacts. As such these issues remain a limitation of non-invasive motion capture systems whether portable and not.

In terms of outcomes, all studies reported kinematics outputs, angles and/or angular velocities, but only 3 studies reported kinetics measures, joint forces and moments. The most commonly used output was the lumbar spine angle, reported as lumbar angle over time, mean and/or maximum and minimum values or in the form of ROM. Outputs were mostly described in the sagittal plane of movement and on a few occasions included measures of the cervical and thoracic spinal segments (Table 3). All outputs were obtained from data processing, including filtering, calibration, integration, mathematical calculations, that took place after the experiments thus precluding real-time biofeedback. Complex data handling algorithms mean long processing time, necessitating specialised hardware and expertise for data interpretation. This is another area that will need further work if data are to be used for clinical assessment and self-management requiring real-time feedback and fast data visualisation of clinically meaningful outputs. Real-time graphical representation of spine motion was only used by Goodvin et al. (2006). Their interface, however, need improvements to be more user friendly and clinically focused.

Overall, all the systems described in the reviewed papers are suitable to study spine movement, however the choice of the system will be dictated by the scope of the study to be conducted. Systems that require PCs and are characterised by the presence of wires are not the best candidate for every-day and prolonged use. In those cases the use of portable data storages will be advocated as described in 8 studies (Boocock et al., 1994, Consmüller et al., 2012, Morlock et al., 2000, Nevins et al., 2002, Thoumie et al., 1998, Van Hoof et al., 2012, Wong and Wong, 2008a, Wong and Wong, 2008b). However, miniaturisation and weight of the portable data logger still need to be optimised. Moreover, if spinal loading is to be measured in the field, in addition to sensors on the spine, sensors on the lower limbs will be necessary for the construction of an inverse-dynamic model as well as portable insoles for ground reaction force measurements as performed by Morlock et al., 2000, Khurelbaatar et al., 2015. If the aim of the study is to provide feedback as a mean to deliver movement intervention and retraining, a visualisation tool similar to the one described by Goodvin et al. (2006) will be needed. The systems described, as mentioned above, where mainly used for validation purposes and their capability in terms of diagnostic in LBP and other musculoskeletal conditions needs to be further explored. Although only used to quantify motion patterns differences, some studies support the use of wearable systems with LBP groups (Dunk and Callaghan, 2010, Lee et al., 2011a, Lee et al., 2011b, Morlock et al., 2000, Van Hoof et al., 2012). Worth mentioning, in this regard, it is that in clinical population, such as LBP, assessment of the spine alongside other body segments may help the diagnostic process as compensation in pain free segments may occur, so the use of a whole body portable sensor system may be required (McGregor and Hukins, 2009, Muller et al., 2015).

There are limitations to be considered when interpreting the findings of this review. Only articles published in English were included posing a language bias to article selection; the quality appraisal was performed based on the checklist developed for the study aim, a standardised tool was not found as study quality was not reported in similar reviews. Finally, the review findings are limited to the articles identified by the set search strategy.

In conclusion, this review shows the use of wearable systems to quantify spine kinematics and kinetics to date. Wearable systems represent valid tools to track spine movement, however, their use is still confined to research studies and at a preliminary stage of development and use. Data logging and processing, systems design and fixation are areas to be improved to fully exploit the wide applicability of wearable technologies overcoming grounded motion capture systems in terms of long-term monitoring and real-life assessment.

Conflict of interest statement

Nothing to declare.

Acknowledgements

This study was funded by the UK Centre for Musculoskeletal Health and Work funded by the Medical Research Council and Arthritis UK.

References

Alonge F., Cucco E., D’Ippolito F., Pulizzotto A. The use of accelerometers and gyroscopes to estimate hip and knee angles on gait analysis. Sensors. 2014;14:8430–8446. [Europe PMC free article] [Abstract] [Google Scholar]
Bartalesi, R., Lorussi, F., De Rossi, D., Tesconi, M., Tognetti, A., 2010. Wearable monitoring of lumbar spine curvature by inertial and e-textile sensory fusion. In: Annual International Conference of the IEEE Engineering in Medicine and Biology, Buenos Aires, pp. 6373–6376. [Abstract]
Baten, C.T.M., Oosterhoff, P., Kingma, I., Veltink, P.H., Hermens, H.J., 1996. Inertial sensing in ambulatory back load estimation. In: Proceedings of 18th Annual International Conference of the IEEE Engineering in Medicine and Biology Society, Amsterdam, pp. 497–498.
Bonato P. Wearable sensors and systems. IEEE Eng. Med. Biol. Mag. 2010;29:25–36. [Abstract] [Google Scholar]
Boocock M.G., Jackson J.A., Burton A.K., Tillotson K.M. Continuous measurement of lumbar posture using flexible electrogoniometers. Ergonomics. 1994;37:175–185. [Abstract] [Google Scholar]
Clays E., De Bacquer D., Leynen F., Kornitzer M., Kittel F De, Backer G. The impact of psychosocial factors on low back pain: longitudinal results from the Belstress study. Spine (Phila Pa 1976) 2007;32:262–268. [Abstract] [Google Scholar]
Charry, E., Umer, M., Taylor, S., 2011. Design and validation of an ambulatory inertial system for 3-D measurements of low back movement. In: 2011 Seventh International Conference on Intelligent Sensors, Sensor Networks and Information Processing, Adelaide, pp. 58–63.
Chhikara, A., McGregor, A.H., Hadjilucas, L., Bello F., Rice, A.S., 2010. Quantitative assessment of the motion of the lumbar spine and pelvis with wearable inertial sensors. In: 2010 International Conference on Body Sensor Networks, Singapore, pp. 9–15.
Consmüller T., Rohlmann A., Weinland D., Druschel C., Duda G.N., Taylor W.R. Comparative evaluation of a novel measurement tool to assess lumbar spine posture and range of motion. Eur. Spine J. 2012;21:2170–2180. [Europe PMC free article] [Abstract] [Google Scholar]
Dagenais S., Caro J., Haldeman S. A systematic review of low back pain cost of illness studies in the United States and internationally. Spine J. 2008;8:8–20. [Abstract] [Google Scholar]
Dankaerts W., O'Sullivan P., Burnett A., Straker L. Differences in sitting postures are associated with nonspecific chronic low back pain disorders when patients are subclassified. Spine (Phila Pa 1976) 2006;31:698–704. [Abstract] [Google Scholar]
Dobson F., Morris M.E., Baker R., Graham H.K. Gait classification in children with cerebral palsy: A systematic review. Gait Posture. 2007;25:140–152. [Abstract] [Google Scholar]
Dunk N.M., Callaghan J.P. Lumbar spine movement patterns during prolonged sitting differentiate low back pain developers from matched asymptomatic controls. Work. 2010;35:3–14. [Abstract] [Google Scholar]
Faber G.S., Chang C.C., Kingma I., van Dennerlein J.T., Dieën J.H. Estimating 3D L5/S1 moments and ground reaction forces during trunk bending using a full-body ambulatory inertial motion capture system. J. Biomech. 2016;49:904–912. [Abstract] [Google Scholar]
Fong D.T.P., Chan Y.Y. The use of wearable inertial motion sensors in human lower limb biomechanics studies: a systematic review. Sensors. 2010;10:11556–11565. [Europe PMC free article] [Abstract] [Google Scholar]
Goodvin C., Park E.J., Huang K., Sakaki K. Development of a real-time three-dimensional spinal motion measurement system for clinical practice. Med. Biol. Eng. Comput. 2006;44:1061–1075. [Abstract] [Google Scholar]
Ha T.H., Saber-Sheikh K., Moore A.P., Jones M.P. Measurement of lumbar spine range of movement and coupled motion using inertial sensors - a protocol validity study. Manual Therapy. 2013;18:87–91. [Abstract] [Google Scholar]
Hernandez A., Gross K., Gombatto S. Differences in lumbar spine and lower extremity kinematics during a step down functional task in people with and people without low back pain. Clin. Biomech. (Bristol, Avon) 2017;47:46–52. [Abstract] [Google Scholar]
Hoy D., March L., Brooks P., Blyth F., Woolf A., Bain C., Williams G., Smith E., Vos T., Barendregt J., Murray C., Burstein R., Buchbinder R. The global burden of low back pain: estimates from the Global Burden of Disease 2010 study. Ann. Rheum. Dis. 2014;73:968–974. [Abstract] [Google Scholar]
Hubble R.P., Naughton G.A., Silburn P.A., Cole M.H. Wearable sensor use for assessing standing balance and walking stability in people with Parkinson’s disease: a systematic review. PLoS One. 2015;10(4):e0123705. [Europe PMC free article] [Abstract] [Google Scholar]
Kerr, M.S., Frank, J.W., Shannon, H.S., Norman, R.W., Wells, R.P., Neumann, W.P., Bombardier, C., Ontario Universities Back Pain Study Group, 2001. Biomechanical and psychosocial risk factors for low back pain at work. Am. J. Public Health 91, 1069–1075. [Abstract]
Khurelbaatar T., Kim K., Lee S., Kim Y.H. Consistent accuracy in whole-body joint kinetics during gait using wearable inertial motion sensors and in-shoe pressure sensors. Gait Posture. 2015;42:65–69. [Abstract] [Google Scholar]
Leardini A., Chiari L., Della Croce U., Cappozzo A. Human movement analysis using stereophotogrammetry: Part 3. Soft tissue artifact assessment and compensation. Gait Posture. 2005;21:212–225. [Abstract] [Google Scholar]
Lee, J.K., Desmoulin, G.T., Khan, A.H., Park, E.J., 2011a. A portable inertial sensing-based spinal motion measurement system for low back pain assessment. In: Annual International Conference of the IEEE Engineering in Medicine and Biology Society, Boston, pp. 4737–4740. [Abstract]
Lee J.K., Desmoulin G.T., Khan A.H., Park E.J. Comparison of 3D spinal motions during stair-climbing between individuals with and without low back pain. Gait Posture. 2011;2011(34):222–226. [Abstract] [Google Scholar]
Lee J.K., Park E.J. 3D spinal motion analysis during staircase walking using an ambulatory inertial and magnetic sensing system. Med. Biol. Eng. Comput. 2011;49(7):755–764. [Abstract] [Google Scholar]
McGregor A.H., Hukins D.W. Lower limb involvement in spinal function and low back pain. J. Back Musculoskelet. Rehabil. 2009;22:219–222. [Abstract] [Google Scholar]
Mitchell T., O'Sullivan P.B., Burnett A.F., Straker L., Smith A. Regional differences in lumbar spinal posture and the influence of low back pain. BMC Musculoskelet. Disord. 2008;9:152. [Europe PMC free article] [Abstract] [Google Scholar]
Morlock M.M., Bonin V., Deuretzbacher G., Müller G., Honl M., Schneider E. Determination of the in vivo loading of the lumbar spine with a new approach directly at the workplace–first results for nurses. Clin. Biomech. (Bristol, Avon) 2000;15:549–558. [Abstract] [Google Scholar]
Muller R., Ertelt T., Blickhan R. Low back pain affects trunk as well as lower limb movements during walking and running. J. Biomech. 2015;48:1009–1014. [Abstract] [Google Scholar]
Needham R., Stebbins J., Chockalingam N. Three-dimensional kinematics of the lumbar spine during gait using marker-based systems: a systematic review. J. Med. Eng. Technol. 2016;40:72–85. [Abstract] [Google Scholar]
Nevins, R.J., Durdle, N.G., Raso, V.J., 2002. A posture monitoring system using accelerometers. In: IEEE CCECE2002, Canadian Conference on Electrical and Computer Engineering Conference Proceedings, pp. 1087–1092.
O’Sullivan K., O’Sullivan L., Campbell A., O’Sullivan P., Dankaerts W. Towards monitoring lumbo-pelvic posture in real-life situations: Concurrent validity of a novel posture monitor and a traditional laboratory-based motion analysis system. Manual Ther. 2012;17:77–83. [Abstract] [Google Scholar]
Ratcliffe E., Pickering S., McLean S., Lewis J. Is there a relationship between subacromial impingement syndrome and scapular orientation? A systematic review. Br. J. Sports Med. 2014;48:1251–1256. [Abstract] [Google Scholar]
Shull P.B., Jirattigalachote W., Hunt M.A., Cutkosky M.R., Delp S.L. Quantified self and human movement: A review on the clinical impact of wearable sensing and feedback for gait analysis and intervention. Gait Posture. 2014;40:11–19. [Abstract] [Google Scholar]
Thoumie P., Drape J.L., Aymard C., Bedoiseau M. Effects of a lumbar support on spine posture and motion assessed by electrogoniometer and continuous recording. Clin. Biomech.(Bristol, Avon) 1998;13:18–26. [Abstract] [Google Scholar]
Van Dillen L.R., Gombatto S.P., Collins D.R., Engsberg J.R., Sahrmann S.A. Symmetry of timing of hip and lumbopelvic rotation motion in 2 different subgroups of people with low back pain. Arch. Phys. Med. Rehabil. 2007;88:351–360. [Abstract] [Google Scholar]
Van Hoof W., Volkaerts K., O'Sullivan K., Verschueren S., Dankaerts W. Comparing lower lumbar kinematics in cyclists with low back pain (flexion pattern) versus asymptomatic controls-field study using a wireless posture monitoring system. Manual Ther. 2012;17:312–317. [Abstract] [Google Scholar]
Walgaard W., van Faber G.S., van Lummel R.C., Dieën J.H., Kingma I. The validity of assessing temporal events, sub-phases and trunk kinematics of the sit-to-walk movement in older adults using a single inertial sensor. J. Biomech. 2016;49:1933–1937. [Abstract] [Google Scholar]
Wong W.Y., Wong M.S. Smart garment for trunk posture monitoring: A preliminary study. Scoliosis. 2008;2008(3):7. [Europe PMC free article] [Abstract] [Google Scholar]
Wong W.Y., Wong M.S. Trunk posture monitoring with inertial sensors. Eur. Spine J. 2008;17:743–753. [Europe PMC free article] [Abstract] [Google Scholar]
Yang S., Li Q. Inertial sensor-based methods in walking speed estimation: a systematic review. Sensors. 2012;12:6102–6116. [Europe PMC free article] [Abstract] [Google Scholar]
Yang C.C., Hsu Y.L. A review of accelerometry-based wearable motion detectors for physical activity monitoring. Sensors. 2010;10:7772–7788. [Europe PMC free article] [Abstract] [Google Scholar]

Citations & impact 


Impact metrics

Jump to Citations

Citations of article over time

Alternative metrics

Altmetric item for https://www.altmetric.com/details/28168323
Altmetric
Discover the attention surrounding your research
https://www.altmetric.com/details/28168323

Article citations


Go to all (48) article citations