Calculation of the time resolution of the J-PET tomograph using kernel density estimation

Phys Med Biol. 2017 Jun 21;62(12):5076-5097. doi: 10.1088/1361-6560/aa7005. Epub 2017 Apr 28.

Abstract

In this paper we estimate the time resolution of the J-PET scanner built from plastic scintillators. We incorporate the method of signal processing using the Tikhonov regularization framework and the kernel density estimation method. We obtain simple, closed-form analytical formulae for time resolution. The proposed method is validated using signals registered by means of the single detection unit of the J-PET tomograph built from a 30 cm long plastic scintillator strip. It is shown that the experimental and theoretical results obtained for the J-PET scanner equipped with vacuum tube photomultipliers are consistent.

MeSH terms

  • Equipment Design
  • Humans
  • Plastics
  • Positron-Emission Tomography / instrumentation*
  • Scintillation Counting / instrumentation*
  • Time Factors