Preclinical evaluation of thin convex probe endobronchial ultrasound-guided transbronchial needle aspiration for intrapulmonary lesions

Transl Lung Cancer Res. 2022 Jul;11(7):1292-1301. doi: 10.21037/tlcr-22-120.

Abstract

Background: Conventional flexible bronchoscopy has not achieved the high diagnostic yield for intrapulmonary lesions as seen with image-guided transthoracic biopsy. A thin convex probe endobronchial ultrasound bronchoscope (TCP-EBUS) with a 5.9-mm tip was designed to improve peripheral access over conventional EBUS bronchoscopes to facilitate real-time sampling of intrapulmonary lesions under ultrasound guidance.

Methods: TCP-EBUS was inserted into the distal airways of ex-vivo human lungs to assess bronchial accessibility relative to clinically available bronchoscopes. The short- (≤1 h) and medium-term (≤10 d) safety of TCP-EBUS insertion and EBUS-guided transbronchial needle aspiration (TBNA) using a 25-gauge needle were evaluated physiologically and radiologically in live pigs. TCP-EBUS-guided TBNA feasibility was assessed in-vivo with pig intrapulmonary pseudo-tumors and ex-vivo with resected human lung cancer specimens.

Results: For bronchial accessibility, TCP-EBUS demonstrated greater reach than the 6.6-mm convex probe endobronchial ultrasound (CP-EBUS) in all bronchi, as well as surpassed a 5.5-mm conventional bronchoscope in 63% (131/209) and a 4.8-mm conventional bronchoscope in 27% (57/209) of assessed bronchi. The median bronchial generation and the mean diameter of bronchi TCP-EBUS reached was 4 (range, 3-7) and 3.3±0.7 mm, respectively. No major complications related to TCP-EBUS-guided TBNA in distal airways were observed in the live pigs. Scattered mucosal erythema of the bronchial walls was observed immediately after TCP-EBUS insertion; this self-resolved by day 10. TCP-EBUS could successfully reach and visualize intrapulmonary targets via ultrasound, with no difficulty in needle deployment or sampling.

Conclusions: TCP-EBUS has the potential to facilitate safe real-time transbronchial sampling of intrapulmonary lesions in the central and middle lung fields.

Keywords: Convex probe endobronchial ultrasound (CP-EBUS); endobronchial ultrasound-guided transbronchial needle aspiration (EBUS-TBNA); intrapulmonary lesions; real-time biopsy.