Abstract
Background
Although Doppler echocardiography has been shown to be accurate in assessing stenotic orifice areas in native valves, its accuracy in evaluating the prosthetic valve orifice area remains undetermined.Methods and results
Doppler-estimated valve areas were studied for their agreement with catheter-derived Gorlin effective orifice areas and their flow dependence in five sizes (19/20-27 mm) of St. Jude, Medtronic-Hall, and Hancock aortic valves using a pulsatile flow model. Doppler areas were calculated three ways: using the standard continuity equation; using its simplified modification (peak flow/peak velocity); and using the Gorlin equation with Doppler pressure gradients. The results were compared with Gorlin effective orifice areas derived from direct flow and catheter pressure measurements. Excellent correlation between Gorlin effective orifice areas and the three Doppler approaches was found in all three valve types (r = 0.93-0.99, SEE = 0.07-0.11 cm2). In Medtronic-Hall and Hancock valves, there was only slight underestimation by Doppler (mean difference, 0.003-0.25 cm2). In St. Jude valves, however, all three Doppler methods significantly underestimated effective orifice areas derived from direct flow and pressure measurements (mean difference, 0.40-0.57 cm2) with differences as great as 1.6 cm2. In general, the modified continuity equation calculated the largest Doppler areas. When orifice areas were calculated from the valve geometry using the area determined from the inner valve diameter reduced by the projected area of the opened leaflets, Gorlin effective orifice areas were much closer to the geometric orifice areas than Doppler areas (mean difference, 0.40 +/- 0.31 versus 1.04 +/- 0.20 cm2). In St. Jude and Medtronic-Hall valves, areas calculated by either technique did not show a consistent or clinically significant flow dependence. In Hancock valves, however, areas calculated by both the continuity equation and the Gorlin equation decreased significantly (p less than 0.001) with low flow rates.Conclusions
Doppler echocardiography using either the continuity equation or Gorlin formula allows in vitro calculation of Medtronic-Hall and Hancock effective valve orifice areas but underestimates valve areas in St. Jude valves. This phenomenon is due to localized high velocities in St. Jude valves, which do not reflect the mean velocity distribution across the orifice. Valve areas are flow independent in St. Jude and Medtronic-Hall prostheses but decrease significantly with low flow in Hancock valves, suggesting that bioprosthetic leaflets may not open fully at low flow rates.Full text links
Read article at publisher's site: https://doi.org/10.1161/01.cir.85.6.2275
Read article for free, from open access legal sources, via Unpaywall: https://www.ahajournals.org/doi/pdf/10.1161/01.CIR.85.6.2275
References
Articles referenced by this article (38)
Khan S Baumgartner H DeRobertis M Czer L Maurer G: Flow dependence of Doppler velocities in normal St Jude valves. JAm Coil Cardiol 1990;15(suppl I):140A
Bileaflet, tilting disc and porcine aortic valve substitutes: in vivo hydrodynamic characteristics.
J Am Coll Cardiol, (2 Pt 1):321-327 1984
MED: 6693620
Title not supplied
Preliminary results in mitral valve replacement with the St Jude Medical prosthesis: Comparison with the Bjork-Sheily valve
Circulation, II-203- 1984
Differences in Hancock and Carpentier-Edwards porcine xenograft aortic valve hemodynamics: Effect of valve size
Circulation 1990
Gorlin R Gorlin SG: Hydraulic formula for the calculation of area of the stenotic mitral valve other cardiac valves and central circulatory shunts. Am Heart J 195 1;41:1-29
Title not supplied
Determination of the stenotic aortic valve area in adults using Doppler echocardiography.
J Am Coll Cardiol, (3):509-517 1986
MED: 3950230
Title not supplied
Prediction of the severity of aortic stenosis by Doppler aortic valve area determination: prospective Doppler-catheterization correlation in 100 patients.
J Am Coll Cardiol, (6):1227-1234 1988
MED: 3366997
Show 10 more references (10 of 38)
Citations & impact
Impact metrics
Citations of article over time
Article citations
Invasive versus echocardiographic gradients in degenerated surgical aortic valve prostheses: A multicenter study.
JTCVS Open, 7:51-60, 17 Jul 2021
Cited by: 1 article | PMID: 36003692 | PMCID: PMC9390586
Comparison of Transvalvular Aortic Mean Gradients Obtained by Intraprocedural Echocardiography and Invasive Measurement in Balloon and Self-Expanding Transcatheter Valves.
J Am Heart Assoc, 10(19):e021014, 29 Sep 2021
Cited by: 12 articles | PMID: 34585593 | PMCID: PMC8649128
Numerical and in-vitro experimental assessment of the performance of a novel designed expanded-polytetrafluoroethylene stentless bi-leaflet valve for aortic valve replacement.
PLoS One, 14(1):e0210780, 30 Jan 2019
Cited by: 4 articles | PMID: 30699210 | PMCID: PMC6353165
Estimating the irreversible pressure drop across a stenosis by quantifying turbulence production using 4D Flow MRI.
Sci Rep, 7:46618, 20 Apr 2017
Cited by: 23 articles | PMID: 28425452 | PMCID: PMC5397859
Effective Orifice Area during Exercise in Bileaflet Mechanical Valve Prostheses.
J Am Soc Echocardiogr, 30(4):404-413, 31 Dec 2016
Cited by: 1 article | PMID: 28049599
Go to all (31) article citations
Similar Articles
To arrive at the top five similar articles we use a word-weighted algorithm to compare words from the Title and Abstract of each citation.
Effect of prosthetic aortic valve design on the Doppler-catheter gradient correlation: an in vitro study of normal St. Jude, Medtronic-Hall, Starr-Edwards and Hancock valves.
J Am Coll Cardiol, 19(2):324-332, 01 Feb 1992
Cited by: 77 articles | PMID: 1531058
Assessment of effective orifice area of prosthetic aortic valves with Doppler echocardiography: an in vivo and in vitro study.
J Thorac Cardiovasc Surg, 122(2):287-295, 01 Aug 2001
Cited by: 15 articles | PMID: 11479501
A comparison of valve resistance, the continuity equation, and the Gorlin formula against directly observed orifice area in bioprosthetic valves in the mitral position: an in vitro study.
J Heart Valve Dis, 5(2):136-143, 01 Mar 1996
Cited by: 3 articles | PMID: 8665004
Doppler echocardiography assessment of prosthetic heart valves.
Echocardiography, 7(2):97-114, 01 Mar 1990
Cited by: 8 articles | PMID: 10149195
Review