Abstract 


Dental implants are the preferred fixed oral rehabilitation for replacing lost teeth. When peri-implant tissues become inflamed, the removal of plaque accumulating around the implant becomes imperative. Recently, several new strategies have been developed for this purpose, with electrolytic decontamination showing increased potential compared to traditional mechanical strategies. In this in vitro pilot study, we compare the efficacy of an electrolytic decontaminant (Galvosurge®) with an erythritol jet system (PerioFlow®) and two titanium brushes (R-Brush™ and i-Brush™) in removing Pseudomonas aeruginosa PAO1 biofilms from implants. Changes in the implant surface after each approach were also evaluated. Twenty titanium SLA implants were inoculated with P. aeruginosa and then randomly assigned to each treatment group. After treatment, decontamination efficacy was assessed by quantifying colony-forming units (log10 CFU/cm2) from each implant surface. Scanning electron microscopy was used to analyse changes in the implant surface. With the exception of R-Brush, all treatment strategies were similarly effective in removing P. aeruginosa from implants. Major surface changes were observed only in implants treated with titanium brushes. In conclusion, this pilot study suggests that electrolytic decontamination, erythritol-chlorhexidine particle jet system and i-Brush™ brushing have similar performance in removing P. aeruginosa biofilm from dental implants. Further studies are needed to evaluate the removal of more complex biofilms. Titanium brushes caused significant changes to the implant surface, the effects of which need to be evaluated.

Free full text 


Logo of jclinmedLink to Publisher's site
J Clin Med. 2023 Feb; 12(4): 1703.
Published online 2023 Feb 20. https://doi.org/10.3390/jcm12041703
PMCID: PMC9967341
PMID: 36836238

Dental Implant Surface Decontamination and Surface Change of an Electrolytic Method versus Mechanical Approaches: A Pilot In Vitro Study

Mariana Anselmo Assunção, Conceptualization, Writing – original draft, Writing – review & editing,1 João Botelho, Data curation, Writing – original draft, Writing – review & editing,2 Vanessa Machado, Writing – review & editing,2 Luís Proença, Data curation, Writing – review & editing,2 António P. A. Matos,2 José João Mendes, Funding acquisition,2 Lucinda J. Bessa, Methodology, Writing – review & editing,2 Nuno Taveira, Conceptualization, Methodology, Validation, Formal analysis, Investigation, Resources, Data curation, Writing – review & editing, Supervision, Project administration,2,* and Alexandre Santos, Conceptualization, Methodology, Writing – review & editing, Visualization, Supervision2,*
Daniele De Santis, Academic Editor

Associated Data

Data Availability Statement

Abstract

Dental implants are the preferred fixed oral rehabilitation for replacing lost teeth. When peri-implant tissues become inflamed, the removal of plaque accumulating around the implant becomes imperative. Recently, several new strategies have been developed for this purpose, with electrolytic decontamination showing increased potential compared to traditional mechanical strategies. In this in vitro pilot study, we compare the efficacy of an electrolytic decontaminant (Galvosurge®) with an erythritol jet system (PerioFlow®) and two titanium brushes (R-Brush™ and i-Brush™) in removing Pseudomonas aeruginosa PAO1 biofilms from implants. Changes in the implant surface after each approach were also evaluated. Twenty titanium SLA implants were inoculated with P. aeruginosa and then randomly assigned to each treatment group. After treatment, decontamination efficacy was assessed by quantifying colony-forming units (log10 CFU/cm2) from each implant surface. Scanning electron microscopy was used to analyse changes in the implant surface. With the exception of R-Brush, all treatment strategies were similarly effective in removing P. aeruginosa from implants. Major surface changes were observed only in implants treated with titanium brushes. In conclusion, this pilot study suggests that electrolytic decontamination, erythritol-chlorhexidine particle jet system and i-Brush™ brushing have similar performance in removing P. aeruginosa biofilm from dental implants. Further studies are needed to evaluate the removal of more complex biofilms. Titanium brushes caused significant changes to the implant surface, the effects of which need to be evaluated.

Keywords: peri-implantitis, implants, decontamination systems, Pseudomonas aeruginosa, electrolytic decontamination, dental implants

1. Introduction

Dental implants are the preferred fixed oral rehabilitation for replacing lost teeth. With their increasing popularity, peri-implant tissue diseases have also increased, with an estimated prevalence of 19.5% [1]. Peri-implantitis is a condition of the tissues surrounding dental implants characterised by an inflammatory response of the peri-implant connective tissue caused by dysbiotic plaque. This results in progressive loss of supporting bone around the implants [2]. Peri-implantitis has been associated with the formation of a bacterial biofilm on the implant surface consisting mainly of Gram-negative anaerobic bacteria, but opportunistic pathogenic microorganisms such as P. aeruginosa, S. aureus and C. albicans can also colonise the implant surface and lead to implant failure. [3,4,5].

Undiagnosed and untreated peri-implantitis has an irregular clinical course and rate of progression [6]. The absence of the periodontal ligament, supracrestal attachment fibres, or root cementum on implants distinguishes peri-implantitis from periodontitis [2]. However, they do share the presence of dental plaque around the surface, which is responsible for the immune response and the inflammatory response. For this reason, current therapies focus on decontaminating the implant surface. There are two types of approaches to the treatment of peri-implantitis: surgical (resective and regenerative surgery) and non-surgical (mechanical, chemical, antibiotic, laser methods and oral hygiene education) [4]. Non-surgical therapy should always precede surgical therapy to reduce peri-implant tissue inflammation, assess response to antimicrobial therapy, and ensure effective patient hygiene [7,8]. However, non-surgical therapy does not usually resolve peri-implantitis [9]. Surgical treatment of peri-implantitis includes mechanical decontamination of the implant surface by debridement of the bacterial biofilm and removal of excess cement and inflamed tissue, creation of peri-implant anatomy that promotes its decontamination, and, if possible, regeneration of infra-osseous defects [9,10]. The ideal therapeutic approach must focus on both the removal of pathogens and the prevention of recolonisation of the implant surface [11]. However, there is still no method for decontaminating the surface of titanium implants that can be considered better and more effective than other existing methods [12].

Among the peri-implant therapeutic approaches, the most commonly used can be categorised into four groups: mechanical, chemical, antibiotic and laser [4,13]. A number of methods, such as titanium brushing, particle blasting and, more recently, the electrolytic approach, have shown good results in the decontamination of titanium surfaces, with each method having a different effect on the surface structure [13,14].

The aim of this study was to compare the plaque removal efficacy of an electrolytic decontamination method with three mechanical methods and to evaluate their impact on the implant surface. Our main hypotheses were that the electrolytic treatment would outperform the mechanical treatments in plaque removal and that the mechanical interventions would have a greater effect on the surface characteristics of the dental implants than the electrolytic treatment. P. aeruginosa was used as a model plaque bacterium because it adheres to the surface of titanium implants and is often associated with peri-implantitis [5].

2. Materials and Methods

2.1. Study Design

This in vitro pilot study compared the removal of P. aeruginosa plaque from SLA implants (NeoCMIImplant-EB-II active Fixture) using an electrolytic approach (Galvosurge®; Galvosurge, Widnau, Switzerland), an erythritol jet system (PerioFlow®; EMS, Nyon, Switzerland) and two titanium brushes (R-Brush™ and i-Brush™; NeoBiotech, Seoul, Republic of Korea). The negative control group received no treatment. We focused on SLA implants because of their widespread clinical use worldwide [15].

2.2. Sample Size Calculation and Allocation of Interventions

In the absence of studies comparing defined surface cleaning of contaminated implants, we defined a pilot sample of 4 implants per group for a total of 20 implants.

After being placed in an in vitro peri-implantitis replication model and contaminated with P. aeruginosa (Section 2.3), the implants were randomly and equally allocated to each group: C (no treatment); EL (electrolytic method); EJ (erythritol jet system); IB (i-Brush); and RB (R-Brush).

2.3. In Vitro Model of Peri-Implantitis and Surface Contamination with P. aeruginosa

SLA implants were immobilised in acrylic models with a tapered defect in the centre and an access hole for implant insertion, leaving the coronal surface exposed (Figure 1).

An external file that holds a picture, illustration, etc.
Object name is jcm-12-01703-g001.jpg

In vitro model of peri-implantitis and surface contamination with P. aeruginosa. After implant placement in the acrylic plate, the surgical cover screw was inserted. The implants were then inverted and immersed in a P. aeruginosa inoculum and incubated at 37 °C for 24 h to allow biofilm formation. The biofilms formed on the implants were then washed by immersion in 2 mL PBS 1×. Surgical cover screws were removed prior to treatment group allocation.

A Pseudomonas aeruginosa PAO1 inoculum was then prepared. Briefly, fresh colonies grown for 24 h on Mueller-Hinton agar (HiMedia Laboratories, India) were used to prepare a bacterial suspension in BHI medium (Brain Heart Infusion; VWR, Belgium) with an optical density at 600 nm (OD600 nm) of 0.1, corresponding to approximately 108 CFU/mL. A further 1:100 dilution was prepared and used to inoculate (1.5 mL) the wells of 12-well plates (JetBiofil Europe, Spain) already containing the acrylics in the inverted position (implants down), as shown in Figure 1. The plates were incubated at 37 °C for 24 h to allow biofilm formation on the implants. The acrylics were then removed and transferred in the same position to a new well containing 2 mL of PBS 1× (Phosphate-Buffered Saline; Fisher Scientific, Oxford, UK) so that the biofilms formed on the surface of the implants were washed once. The acrylic was then positioned with the implants side-up and ready for the subsequent specific treatment.

2.4. Plaque Removal Interventions

  • Electrolytic decontamination [Figure 2a]: Electrolytic decontamination device using galvanic currents (Galvosurge; Galvosurge Dental AG, Widnau, Switzerland), used according to the manufacturer’s instructions. The head of the device is inserted into the implant and held in place throughout the cleaning process. The Galvosurge cleaning solution is pumped in, and a galvanic current forms hydrogen bubbles on the implant surface to release the adherent biofilm. The unit automatically stops working when it reaches 100% (2 min).

    An external file that holds a picture, illustration, etc.
Object name is jcm-12-01703-g002.jpg

    Instruments used to remove plaque from the implants: (a) Galvosurge®; (b) Perioflow®; (c) i-Brush; (d) R-Brush.

  • Air abrasion with erythritol powder [Figure 2b]: An erythritol-based air-powder abrasive device with a subgingival plastic nozzle (Perioflow; EMS, Nyon, Switzerland), used according to the manufacturer’s recommendations, with the “liquid” and “power” settings set to the maximum (2 min).

  • Mechanical debridement with i-Brush™ [Figure 2c]: Titanium brushes with stainless bristles (i-Brush; NeoBiotech, Seoul, Republic of Korea) used with irrigation at 10,000 rpm (2 min).

  • Mechanical debridement using R-Brush™ [Figure 2d]: Rotary titanium brushes (5.6/6.6 in diameter; R-Brush; NeoBiotech, Seoul, Republic of Korea) were used at a rotational speed of 5000 rpm under irrigation with water (2 min).

2.5. Biofilm Quantification

Each implant, either untreated (control) or treated, was carefully removed from the acrylic and placed in the well of a new 12-well plate containing 2 mL PBS 1×. Each 12-well plate was sealed with parafilm and placed in an ultrasonic bath (TPC Dentsonic UC-1000 10L Ultrasonic Cleaner) for 20 min to dislodge and disperse the biofilm from the implants. Then, from each well containing the dispersed biofilm in PBS 1×, 100 μL was used (after vigorous up-down pipetting for better homogenisation) to perform serial 1:10 dilutions. Controls were diluted from 10−1 to 10−6, and treated conditions were diluted from 10−1 to 10−4. Then 100 μL of the last four dilutions (10−3 to 10−6 for the control conditions and 10−1 to 10−4 for the treated conditions) were plated on Mueller-Hinton agar plates and incubated overnight at 37 °C. The next day, colony-forming units (CFU) were counted on agar plates containing between 30 and 300 colonies. CFU/cm2 values were then calculated for each condition. The area (cm2) of the implant where the biofilm was formed was calculated using the following formula: A = 2.π.r.h, where ‘r’ is the radius of the implant and ‘h’ is the height immersed in the inoculum. CFU/cm2 were converted to log10 CFU/cm2, and the mean and standard deviation were calculated for each study group.

2.6. Scanning Electron Microscopy

The samples were observed using the following parameters: 350× magnification, 40 mm working distance (WD), and 30 kV high voltage. Images were processed using the GNU Image Manipulation Program—gimp-2.10.

2.7. Statistical Analysis

Differences in bacterial colony counts (log10 CFU/cm2) between control and treatment groups were evaluated by one-way ANOVA with Tukey’s multiple comparison test. Hypothesis tests were two-tailed, and values of p < 0.05 were considered significant.

3. Results

3.1. Quantitative Analysis of Implant-Adhered Bacteria

With the exception of the R-Brush treatment, the number of P. aeruginosa cells removed from the SLA implants after the different types of treatment was significantly lower compared to the control implants (no treatment) (Figure 3, Table 1 and Table 2). Consistent with this, the RB group had the highest mean log10 CFU/cm2 of all treatment groups. On the other hand, the EJ (erythritol jet) group had the lowest mean CFU/cm2 values (Figure 3 and Table 1). These data suggest that air polishing with an erythritol jet is the best method for decontaminating implants and that the titanium R-Brush treatment is not effective for this purpose.

An external file that holds a picture, illustration, etc.
Object name is jcm-12-01703-g003.jpg

P. aeruginosa PAO1 CFU counts (log10 CFU/cm2) obtained—from the no treatment control group (C) and for the treatment groups (electrolytic decontamination (EL); erythritol jet (EJ); i-Brush™ (IB) and R-Brush™ (RB)). Mean and standard deviation are presented.

Table 1

Descriptive results (log10 CFU/cm2) for each group.

PCELEJIBRB
Mean (SD)8.39 (0.36)6.46 (1.07)5. 86 (1.32)6.32 (0.38)6.74 (0.33)
SE0.180.540.660.190.16
95% CI7.83–8.964.76–8.163.76–7.955.72–6.936.22–7.26
Minimum7.885.724.095.856.39
Maximum8.698.007.276.777.13

Table 2

Multiple comparisons between the groups (p values are shown).

CELEJIBRB
C-0.0280.0040.0180.070
EL--0.8240.9990.987
EJ---0.9200.549
IB----0.947
RB-----

3.2. Qualitative Evaluation of Treated Implants by Scanning Electron Microscopy

The surface of the four implants in each group was evaluated by scanning electron microscopy, and the valley area between the coils was observed in two different areas of each implant, giving a total of eight images per group (Figure 4). There was a noticeable change in the surface of the implants treated with i-Brush™ and R-Brush™. In addition, when comparing these two brushes, there appears to be a greater change on the surfaces instrumented with i-Brush™. In the EL and EJ groups, no significant changes were observed on the different surfaces compared to the control group.

An external file that holds a picture, illustration, etc.
Object name is jcm-12-01703-g004.jpg

Surface analysis of treated implants by scanning electron microscopy: (a) control group; (b) EL: electrolytic decontamination; (c) EJ: erythritol jet; (d) IB: i-Brush™; (e) RB: R-Brush™. Representative images are shown.

4. Discussion

This study showed that electrolytic decontamination, erythritol jet and i-Brush titanium brushes significantly reduced the bacterial load on titanium implant surfaces. However, none of the methods achieved complete decontamination of the implants. Lower bacterial counts (log10 CFU/cm2) were observed on implants treated with the air polishing system, and higher bacterial counts were observed on implants treated with R-Brush, but the differences between methods did not reach statistical significance, indicating that all decontamination strategies except R-Brush performed similarly. As a result, the hypothesis of this study that the electrolytic treatment would be better than the mechanical treatments in biofilm removal is rejected. To our knowledge, there are no in vitro studies performed under the same conditions comparing the effectiveness of the decontamination methods evaluated in this study.

SEM is commonly used for qualitative analysis of implant surfaces because of its ability to provide information about the surface topography of the samples [11,13,16,17,18,19,20,21,22,23,24,25]. No significant changes were observed on the implant surfaces in the EL and EJ groups, as observed by SEM. In contrast, implants instrumented with the titanium brushes (groups IB and RB) showed significant surface changes in the form of scratches compatible with the brush bristles. These results confirm the second hypothesis of this study that mechanical interventions have a greater effect on the surface characteristics of dental implants than electrolytic treatment.

Mechanical debridement of biofilm using titanium brushes has shown some efficacy in previous studies, i.e., a reasonable reduction in the bacterial count of the surface under investigation [17,18,19,25,26]. However, brushing did not show superior efficacy compared to other existing methods, and the authors also reported significant changes in the titanium surfaces caused by brushing, using qualitative analysis by SEM. Our results confirm the previously mentioned studies, i.e., the i-Brush showed some decontamination efficacy without evidence of superiority compared to the other methods used and caused remarkable changes in the surface of the implants. The relatively poor decontamination performance of the R-Brush observed in the current study is likely related to its low rotation speed (5000 rpm, half that of the i-Brush).

The particle jet system is a widely studied method, especially when using sodium bicarbonate particles [22,24,27,28] and glycine [13,14,21,22,23,25,27,28,29,30]. It has been found that when smaller particle sizes such as glycine (25 μm) or erythritol (14 μm) are used, the surface changes are not significant compared to sodium bicarbonate (40-65 μm), which has larger particle sizes and consequently higher abrasive power [22]. In the present study, we found that the erythritol-chlorhexidine particle jet system (14 μm) had no significant effect on the surface of the SLA implants. These results are consistent with previous studies showing that this treatment does not cause significant changes to the surface of implants [22].

Recently, a new electrolytic approach has been developed, which consists of applying a galvanic current to the implant, resulting in the formation of hydrogen bubbles on its surface, thus releasing the adherent biofilm [13]. Electrochemical treatment of dental implants combines the antibacterial action of bactericides with the direct oxidation of bacterial enzymes and proteins, resulting in a low free concentration of harmful chemicals, allowing the surface of an implant to be cleaned without altering its microtopography and without affecting its physical properties [13,31,32,33]. In these studies, electrolytic decontamination outperformed glycine jetting, and only electrolytic decontamination was able to completely decontaminate the implant surfaces [13]. In the present study, electrolytic decontamination did not show high decontamination performance, with higher CFU/cm2 values than the erythritol jet method. The difference in results between the present study and previous studies may be due to the use of different particles in the jetting system and the difference in the bacterial system used, i.e., P. aeruginosa biofilm vs saliva-derived biofilm. The bacteria present in the saliva-derived biofilm were not characterised, and the salivary microbiome is usually highly heterogeneous [34], making it impossible to compare the results. However, we have used a robust and strongly adherent biofilm model of P. aeruginosa, which can be considered more difficult to remove, thus contributing to the poorer results obtained in this study [35]. P. aeruginosa is a gram-negative bacterium that grows easily under a wide range of conditions and temperatures and has a high ability to adhere to surfaces [35]. In addition, P. aeruginosa is commonly found on dental implants in patients with peri-implantitis [5]. For these reasons, P. aeruginosa was chosen as the model bacterium for this research. On the other hand, the SEM evaluation of the surface of the implants treated by the electrolytic decontamination method confirms previous results [13,33], i.e., the electrolytic decontamination system does not cause significant changes in the surface of the implants.

This in vitro study has limitations such as the use of implants with different geometries, different areas exposed to contamination and decontamination, non-uniformity of the biofilm formed on the implant surface due to the characteristics of P. aeruginosa biofilms, and the use of monospecies biofilms. Studies using multi-species biofilms, which better mimic the complex biofilm found in the oral cavity, are needed to better determine the usefulness of the methods analysed in this study for implant decontamination.

5. Conclusions

This pilot study suggests that electrolytic treatment, erythritol-chlorhexidine particle jetting and i-Brush™ brushing have similar performance in removing P. aeruginosa biofilm from dental implants. Further studies are needed to evaluate the removal of more complex biofilms, which represent dental plaque. Titanium brushes caused significant changes to the implant surface, the effects of which need to be evaluated.

Acknowledgments

We thank NeoBiotech and EMS for their generous material donations that allowed us to conduct this study.

Funding Statement

This research received no external funding.

Author Contributions

Conceptualization, N.T., M.A.A., A.P.A.M. and A.S.; methodology, M.A.A., A.S., N.T., A.P.A.M., L.P. and L.J.B.; formal analysis, M.A.A., A.P.A.M., L.J.B., L.P., N.T. and A.S.; investigation, M.A.A., A.P.A.M., A.S. and L.J.B.; resources, J.J.M.; data curation, M.A.A., A.S., A.P.A.M., J.B. and V.M.; writing—original draft preparation, M.A.A.; writing—review and editing, N.T., A.S., A.P.A.M., J.J.M., J.B., V.M. and L.P.; visualization, A.P.A.M. and M.A.A.; supervision, N.T., L.P., A.P.A.M. and A.S.; funding acquisition, J.J.M. All authors have read and agreed to the published version of the manuscript.

Institutional Review Board Statement

Not applicable.

Informed Consent Statement

Not applicable.

Data Availability Statement

Not applicable.

Conflicts of Interest

The authors declare no conflict of interest.

Footnotes

Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content.

References

1. Diaz P., Gonzalo E., Villagra L.J.G., Miegimolle B., Suarez M.J. What Is the Prevalence of Peri-Implantitis? A Systematic Review and Meta-Analysis. BMC Oral Health. 2022;22:449. 10.1186/s12903-022-02493-8. [Europe PMC free article] [Abstract] [CrossRef] [Google Scholar]
2. Berglundh T., Jepsen S., Stadlinger B., Terheyden H. Peri-Implantitis and Its Prevention. Clin. Oral Implant. Res. 2019;30:150–155. 10.1111/clr.13401. [Abstract] [CrossRef] [Google Scholar]
3. Albertini M., López-Cerero L., O’Sullivan M.G., Chereguini C.F., Ballesta S., Ríos V., Herrero-Climent M., Bullón P. Assessment of Periodontal and Opportunistic Flora in Patients with Peri-Implantitis. Clin. Oral Implant. Res. 2015;26:937–941. 10.1111/clr.12387. [Abstract] [CrossRef] [Google Scholar]
4. Rokaya D., Srimaneepong V., Wisitrasameewon W., Humagain M., Thunyakitpisal P. Peri-Implantitis Update: Risk Indicators, Diagnosis, and Treatment. Eur. J. Dent. 2020;14:672–682. 10.1055/s-0040-1715779. [Europe PMC free article] [Abstract] [CrossRef] [Google Scholar]
5. El-Telbany M., El-Sharaki A. Antibacterial and Anti-Biofilm Activity of Silver Nanoparticles on Multi-Drug Resistance Pseudomonas Aeruginosa Isolated from Dental-Implant. J. Oral Biol. Craniofacial Res. 2022;12:199–203. 10.1016/j.jobcr.2021.12.002. [Europe PMC free article] [Abstract] [CrossRef] [Google Scholar]
6. Derks J., Schaller D., Håkansson J., Wennström J.L., Tomasi C., Berglundh T. Peri-Implantitis—Onset and Pattern of Progression. J. Clin. Periodontol. 2016;43:383–388. 10.1111/jcpe.12535. [Abstract] [CrossRef] [Google Scholar]
7. Renvert S., Hirooka H., Polyzois I., Kelekis-Cholakis A., Wang H.-L. Diagnosis and Non-Surgical Treatment of Peri-Implant Diseases and Maintenance Care of Patients with Dental Implants—Consensus Report of Working Group 3. Int. Dent. J. 2019;69:12–17. 10.1111/idj.12490. [Europe PMC free article] [Abstract] [CrossRef] [Google Scholar]
8. Wang C.-W., Renvert S., Wang H.-L. Nonsurgical Treatment of Periimplantitis. Implant. Dent. 2019;28:155–160. 10.1097/ID.0000000000000846. [Abstract] [CrossRef] [Google Scholar]
9. Kormas I., Pedercini C., Pedercini A., Raptopoulos M., Alassy H., Wolff L.F. Peri-Implant Diseases: Diagnosis, Clinical, Histological, Microbiological Characteristics and Treatment Strategies. A Narrative Review. Antibiotics. 2020;9:835. 10.3390/antibiotics9110835. [Europe PMC free article] [Abstract] [CrossRef] [Google Scholar]
10. Khoury F., Keeve P.L., Ramanauskaite A., Schwarz F., Koo K.-T., Sculean A., Romanos G. Surgical Treatment of Peri-Implantitis—Consensus Report of Working Group 4. Int. Dent. J. 2019;69:18–22. 10.1111/idj.12505. [Europe PMC free article] [Abstract] [CrossRef] [Google Scholar]
11. Amate-Fernández P., Figueiredo R., Blanc V., Àlvarez G., León R., Valmaseda-Castellón E. Erythritol-Enriched Powder and Oral Biofilm Regrowth on Dental Implants: An in Vitro Study. Med. Oral Patol. Oral Cir. Bucal. 2021;26:e602–e610. 10.4317/medoral.24622. [Europe PMC free article] [Abstract] [CrossRef] [Google Scholar]
12. Toma S., Brecx M.C., Lasserre J.F. Clinical Evaluation of Three Surgical Modalities in the Treatment of Peri-Implantitis: A Randomized Controlled Clinical Trial. J. Clin. Med. 2019;8:966. 10.3390/jcm8070966. [Europe PMC free article] [Abstract] [CrossRef] [Google Scholar]
13. Ratka C., Weigl P., Henrich D., Koch F., Schlee M., Zipprich H. The Effect of In Vitro Electrolytic Cleaning on Biofilm-Contaminated Implant Surfaces. J. Clin. Med. 2019;8:1397. 10.3390/jcm8091397. [Europe PMC free article] [Abstract] [CrossRef] [Google Scholar]
14. Alovisi M., Carossa M., Mandras N., Roana J., Costalonga M., Cavallo L., Pira E., Putzu M.G., Bosio D., Roato I., et al. Disinfection and Biocompatibility of Titanium Surfaces Treated with Glycine Powder Airflow and Triple Antibiotic Mixture: An In Vitro Study. Materials. 2022;15:4850. 10.3390/ma15144850. [Europe PMC free article] [Abstract] [CrossRef] [Google Scholar]
15. Rupp F., Liang L., Geis-Gerstorfer J., Scheideler L., Hüttig F. Surface Characteristics of Dental Implants: A Review. Dent. Mater. 2018;34:40–57. 10.1016/j.dental.2017.09.007. [Abstract] [CrossRef] [Google Scholar]
16. Vernon-Parry K.D. Scanning Electron Microscopy: An Introduction. III-Vs Rev. 2000;13:40–44. 10.1016/S0961-1290(00)80006-X. [CrossRef] [Google Scholar]
17. Al-Hashedi A.A., Laurenti M., Benhamou V., Tamimi F. Decontamination of Titanium Implants Using Physical Methods. Clin. Oral Implant. Res. 2017;28:1013–1021. 10.1111/clr.12914. [Abstract] [CrossRef] [Google Scholar]
18. Alotaibi M., Moran G., Grufferty B., Renvert S., Polyzois I. The Effect of a Decontamination Protocol on Contaminated Titanium Dental Implant Surfaces with Different Surface Topography in Edentulous Patients. Acta Odontol. Scand. 2019;77:66–75. 10.1080/00016357.2018.1504986. [Abstract] [CrossRef] [Google Scholar]
19. El Chaar E., Almogahwi M., Abdalkader K., Alshehri A., Cruz S., Ricci J. Decontamination of the Infected Implant Surface: A Scanning Electron Microscope Study. Int. J. Periodontics Restor. Dent. 2020;40:395–401. 10.11607/prd.4568. [Abstract] [CrossRef] [Google Scholar]
20. Furtsev T.V., Zeer G.M. Efficiency of Cleaning the Various Types of Dental Implants’ Surfaces (Tiu-Nite, Sla, Rbm) Using the Air-Flow Erythritol Method. J. Int. Dent. Med. Res. 2020;13:448–452. [Google Scholar]
21. Keim D., Nickles K., Dannewitz B., Ratka C., Eickholz P., Petsos H. In Vitro Efficacy of Three Different Implant Surface Decontamination Methods in Three Different Defect Configurations. Clin. Oral Implant. Res. 2019;30:550–558. 10.1111/clr.13441. [Abstract] [CrossRef] [Google Scholar]
22. Matsubara V.H., Leong B.W., Leong M.J.L., Lawrence Z., Becker T., Quaranta A. Cleaning Potential of Different Air Abrasive Powders and Their Impact on Implant Surface Roughness. Clin. Implant. Dent. Relat. Res. 2020;22:96–104. 10.1111/cid.12875. [Abstract] [CrossRef] [Google Scholar]
23. Otsuki M., Wada M., Yamaguchi M., Kawabata S., Maeda Y., Ikebe K. Evaluation of Decontamination Methods of Oral Biofilms Formed on Screw-Shaped, Rough and Machined Surface Implants: An Ex Vivo Study. Int. J. Implant. Dent. 2020;6:18. 10.1186/s40729-020-00212-y. [Europe PMC free article] [Abstract] [CrossRef] [Google Scholar]
24. Passarelli P.C., De Leonardis M., Piccirillo G.B., Desantis V., Papa R., Rella E., Mastandrea Bonaviri G.N., Papi P., Pompa G., Pasquantonio G., et al. The Effectiveness of Chlorhexidine and Air Polishing System in the Treatment of Candida Albicans Infected Dental Implants: An Experimental In Vitro Study. Antibiotics. 2020;9:179. 10.3390/antibiotics9040179. [Europe PMC free article] [Abstract] [CrossRef] [Google Scholar]
25. Sanz-Martín I., Paeng K., Park H., Cha J.-K., Jung U.-W., Sanz M. Significance of Implant Design on the Efficacy of Different Peri-Implantitis Decontamination Protocols. Clin. Oral Investig. 2021;25:3589–3597. 10.1007/s00784-020-03681-y. [Abstract] [CrossRef] [Google Scholar]
26. AlMoharib H.S., Steffensen B., Zoukhri D., Finkelman M., Gyurko R. Efficacy of an Er:YAG Laser in the Decontamination of Dental Implant Surfaces: An in Vitro Study. J. Periodontol. 2021;92:1613–1621. 10.1002/JPER.20-0765. [Abstract] [CrossRef] [Google Scholar]
27. Biazussi B.R., Perrotti V., D’Arcangelo C., Elias C.N., Bianchini M.A., Tumedei M., de Vasconcellos D.K. Evaluation of the Effect of Air Polishing With Different Abrasive Powders on the Roughness of Implant Abutment Surface: An In Vitro Study. J. Oral Implantol. 2019;45:202–206. 10.1563/aaid-joi-D-18-00156. [Abstract] [CrossRef] [Google Scholar]
28. Pranno N., Cristalli M.P., Mengoni F., Sauzullo I., Annibali S., Polimeni A., La Monaca G. Comparison of the Effects of Air-Powder Abrasion, Chemical Decontamination, or Their Combination in Open-Flap Surface Decontamination of Implants Failed for Peri-Implantitis: An Ex Vivo Study. Clin. Oral Investig. 2021;25:2667–2676. 10.1007/s00784-020-03578-w. [Europe PMC free article] [Abstract] [CrossRef] [Google Scholar]
29. Drago L., Del Fabbro M., Bortolin M., Vassena C., De Vecchi E., Taschieri S. Biofilm Removal and Antimicrobial Activity of Two Different Air-Polishing Powders: An In Vitro Study. J. Periodontol. 2014;85:e363–e369. 10.1902/jop.2014.140134. [Abstract] [CrossRef] [Google Scholar]
30. Mensi M., Cochis A., Sordillo A., Uberti F., Rimondini L. Biofilm Removal and Bacterial Re-Colonization Inhibition of a Novel Erythritol/Chlorhexidine Air-Polishing Powder on Titanium Disks. Materials. 2020;11:1510. 10.3390/ma11091510. [Europe PMC free article] [Abstract] [CrossRef] [Google Scholar]
31. Schneider S., Rudolph M., Bause V., Terfort A. Electrochemical Removal of Biofilms from Titanium Dental Implant Surfaces. Bioelectrochemistry. 2018;121:84–94. 10.1016/j.bioelechem.2018.01.008. [Abstract] [CrossRef] [Google Scholar]
32. Gianfreda F., Punzo A., Pistilli V., Bollero P., Cervino G., D’Amico C., Cairo F., Cicciù M. Electrolytic Cleaning and Regenerative Therapy of Peri-Implantitis in the Esthetic Area: A Case Report. Eur. J. Dent. 2022;16:950–956. 10.1055/s-0042-1750773. [Europe PMC free article] [Abstract] [CrossRef] [Google Scholar]
33. Schlee M., Rathe F., Brodbeck U., Ratka C., Weigl P., Zipprich H. Treatment of Peri-Implantitis—Electrolytic Cleaning Versus Mechanical and Electrolytic Cleaning—A Randomized Controlled Clinical Trial—Six-Month Results. J. Clin. Med. 2019;8:1909. 10.3390/jcm8111909. [Europe PMC free article] [Abstract] [CrossRef] [Google Scholar]
34. Ruan X., Luo J., Zhang P., Howell K. The Salivary Microbiome Shows a High Prevalence of Core Bacterial Members yet Variability across Human Populations. NPJ Biofilms Microbiomes. 2022;8:85. 10.1038/s41522-022-00343-7. [Europe PMC free article] [Abstract] [CrossRef] [Google Scholar]
35. LaBauve A.E., Wargo M.J. Growth and Laboratory Maintenance of Pseudomonas Aeruginosa. Curr. Protoc. Microbiol. 2012;25:6E.1.1–6E.1.8. 10.1002/9780471729259.mc06e01s25. [Europe PMC free article] [Abstract] [CrossRef] [Google Scholar]

Articles from Journal of Clinical Medicine are provided here courtesy of Multidisciplinary Digital Publishing Institute (MDPI)

Citations & impact 


Impact metrics

Jump to Citations

Smart citations by scite.ai
Smart citations by scite.ai include citation statements extracted from the full text of the citing article. The number of the statements may be higher than the number of citations provided by EuropePMC if one paper cites another multiple times or lower if scite has not yet processed some of the citing articles.
Explore citation contexts and check if this article has been supported or disputed.
https://scite.ai/reports/10.3390/jcm12041703

Supporting
Mentioning
Contrasting
0
2
0

Article citations

Similar Articles 


To arrive at the top five similar articles we use a word-weighted algorithm to compare words from the Title and Abstract of each citation.