Europe PMC

This website requires cookies, and the limited processing of your personal data in order to function. By using the site you are agreeing to this as outlined in our privacy notice and cookie policy.

Abstract 


Background

Catheter-related bloodstream infections (CRBSIs) are a major cause of morbidity and mortality among hospitalized patients. Different studies suggest that the use of disinfectant caps (DCs) significantly reduces the rate of CRBSIs. The first purpose of this study is to analyze, through an in-vitro-model, the antiseptic effect of DCs produced by two manufacturers; the second aim is to assess potential differences in terms of effectiveness between the two manufacturers' products.

Methods

A know concentration of thirteen different microorganisms was incubated with the sponge drenched in antimicrobial fluid inside DCs and cultured through several assays to investigate the disinfectant effectiveness of some commercially available caps. Disinfectant properties were evaluated under two different conditions: baseline (DCs placed on the needle-free connectors (NFCs) and stress test (DCs directly applied to the catheter hub).

Results

Both manufacturers overcame the basal tests (fourteen different assays). Regarding stress tests: the only significant bacterial load was found for Serratia marcescens (104 CFU/mL in ICU Medical™), both at 90 and 180 minutes after incubation; due to the low load, MDR Acinetobacter baumannii was not considered significant (<103 CFU/mL in BD PureHub™).

Conclusions

Our results confirm what was reported in BD PureHub™ datasheet and add data not previously shown by ICU Medical™. Moreover, no difference was observed between the two manufacturers products: the use of both DCs on NFCs was able to reclaim the catheter lumen. These findings support the routine use of DCs with NFCs, as part of a structured bundle of interventions, to reduce the incidence of CRBSIs.

Free full text 


Logo of infmedLink to Publisher's site
Infez Med. 2023; 31(4): 553–559.
Published online 2023 Dec 1. https://doi.org/10.53854/liim-3104-14
PMCID: PMC10705867
PMID: 38075418

Disinfectant caps in vitro effectiveness

SUMMARY

Background

Catheter-related bloodstream infections (CRBSIs) are a major cause of morbidity and mortality among hospitalized patients. Different studies suggest that the use of disinfectant caps (DCs) significantly reduces the rate of CRBSIs. The first purpose of this study is to analyze, through an in-vitro-model, the antiseptic effect of DCs produced by two manufacturers; the second aim is to assess potential differences in terms of effectiveness between the two manufacturers’ products.

Methods

A know concentration of thirteen different microorganisms was incubated with the sponge drenched in antimicrobial fluid inside DCs and cultured through several assays to investigate the disinfectant effectiveness of some commercially available caps. Disinfectant properties were evaluated under two different conditions: baseline (DCs placed on the needle-free connectors (NFCs) and stress test (DCs directly applied to the catheter hub).

Results

Both manufacturers overcame the basal tests (fourteen different assays). Regarding stress tests: the only significant bacterial load was found for Serratia marcescens (104 CFU/mL in ICU Medical™), both at 90 and 180 minutes after incubation; due to the low load, MDR Acinetobacter baumannii was not considered significant (<103 CFU/mL in BD PureHub™).

Conclusions

Our results confirm what was reported in BD PureHub™ datasheet and add data not previously shown by ICU Medical™. Moreover, no difference was observed between the two manufacturers products: the use of both DCs on NFCs was able to reclaim the catheter lumen. These findings support the routine use of DCs with NFCs, as part of a structured bundle of interventions, to reduce the incidence of CRBSIs.

Keywords: Catheter-related bloodstream infections (CRBSIs), needle-free connectors (NFCs), disinfectant cap (DCs), Port Protector, in vitro model

INTRODUCTION

Catheter-related bloodstream infections (CRBSIs) are clinical conditions characterized by fever and systemic clinical signs. A CRBSI should be diagnosed using paired qualitative (as measured using differential time to positivity-DTP) and/or using paired quantitative (as measured using pour plates) blood cultures from a peripheral vein and from the catheter. For hospitals using DTP, a CRBSI is diagnosed if the same organism is isolated from blood obtained through the catheter hub and from blood obtained from a peripheral vein and the DTP is more than 2 hours (catheter hub culture positive first) [1]. CRBSIs are potentially life-threatening healthcare associated infections, that continue to be one of the most important public health issues throughout the world [2]. Although the reported incidence of CRBSIs is around 0.4 percent, these clinical entities constitute an important cause of hospital-acquired infections associated with elevated morbidity, mortality and increased costs [3, 4].

Researchers have reported that the use of 70% isopropyl alcohol just before needle free connectors (NFCs) is not enough against microbial contamination: most health care organizations use a wipe impregnated with 2% chlorhexidine and 70% isopropyl alcohol, which has been shown to decontaminate NFCs when applied for at least 15 seconds, as recommended by evidence-based guidelines [57]. However, manual disinfection of NFCs is significantly affected by expertise and sometimes lack of healthcare workers compliance [8]. Therefore, as recommended by guidelines, NFCs disinfection should be achieved using a disinfectant cap (DC) [5, 8, 9].

The DC (also known as Disinfecting Port Protector) is a plastic device that contains 70% isopropyl alcohol. It is effective within 10 minutes from application and may be used for up to 7 days. Several studies and a recent meta-analysis suggest that the clinical use of DCs significantly reduced the incidence of CRBSIs when compared with manual disinfection [915].

To the best of our knowledge, there aren’t independent studies aiming to evaluate DCs effectiveness and to point out potential differences between what the manufacturers claim in their datasheets. The first purpose of this study is to analyze, in an in vitro model, the antiseptic effect of the DCs produced by two manufacturers; we evaluated if, in the event of maximum bacterial contamination, the DCs could still succeed in reclaiming the catheter from various micro-organisms. The second purpose was to analyze potential differences in terms of effectiveness between the two manufacturers products.

MATERIALS AND METHODS

From April 2022 to February 2023, the Microbiology Unit of ASST Fatebenefratelli-Sacco, Milan, Italy, conducted a study to investigate the efficacy of some commercially available DCs, under two different conditions: the baseline, recommended by guidelines (application of DC to NFC) and a stress test (placing the DC directly on the catheter hub) [16]. We then compared the antimicrobial capacity of various Port Protectors in these two conditions. In the first phase of the study (basal test), a known concentration (0.5 McFarland) of different micro-organisms was inoculated in the two NFCs applied on a two-lumen catheter (Arrow®, Teleflex; 20cm length 7 FR dual lumen). After bacterial growth was completed (12 hours after inoculation), we used two disinfectant caps belonging to two different brands (BD PureHub™ and ICU Medical™) as shown in Figure 1. The DCs were placed on NFCs as indicated by guidelines [17]. Ten minutes after application (as stated in DCs datasheets), we removed the sponge drenched in antimicrobial fluid inside the DC and cultured it for 90 and 180 minutes after incubation. To confirm proper contamination, also NFCs were cultured. In sample size calculation, we considered a single proportion with an estimated 3% positivity and a 10% accuracy in terms of confidence level 95%. We calculated a sample size of at least 12 cultures, increased to 14 due to an expected 10% drop out. In the second phase, a known concentration (0.5 MF) of different micro-organisms was inoculated in the two hubs of a two-lumen catheter (Arrow®, Teleflex; 20cm length 7 FR dual lumen) as shown in Figure 1.

An external file that holds a picture, illustration, etc.
Object name is 1124-9390_31_4_2023_553-559f1.jpg

A known concentration (0.5 MF) of different microorganisms was inoculated into a two-lumen catheter (A, B). At the end of bacterial growth, disinfectant caps of two main brands (BD PureHub™ and ICU Medical™) were screwed onto each catheter lumen and retained on it for ten minutes (C). After this time, we removed the filter inside the disinfectant caps (D) and cultured it for 90 and 180 minutes after incubation

After bacterial growth was completed (12 hours after inoculation), we used two disinfectant caps of two different brands (BD PureHub™ and ICU Medical™). The DCs were applied directly on the catheter hubs.

Ten minutes after application (as stated in DCs datasheets), we removed the sponge drenched in antimicrobial fluid inside the DC and cultured it for 90 and 180 minutes after incubation.

We decided to define this as “stress test” because the hollow surface of the catheter hub is usually harder to be disinfected than the smooth surface of the NFC.

To confirm proper contamination, distal catheters hubs were cultured.

The MALDI-TOF MS systems (bioMerieux Coroprate) was used for all bacterial identification in both first and second phase.

In the second phase sample size calculation, we considered a single proportion with an estimated 16% positivity and a 10% accuracy in terms of confidence level 95%. We calculated a sample size of at least 52 culture, increased to 58 due to an expected 10% drop out.

According to the literature and in order to increase the sensitivity value of the test, in both baseline and stress test, each microbial strain was cultured both at 90 and 180 minutes after incubation [18]. Only microbial loads above 104 were considered relevant [18].

Thirteen different strains of the most involved microorganisms in the setting of hospital acquired infections (6 Gram-positive bacteria, 6 Gram-negative bacteria, 1 yeast), were tested.

RESULTS

During the baseline test, fourteen different assays were performed: 3 Gram negative bacteria (carbapenem-resistant Klebsiella pneumoniae, multidrug-resistant Pseudomonas aeruginosa, multidrug-resistant Acinetobacter pittii), one the most involved bacteria in soft tissue infections (methicillin-resistant Staphylococcus aureus) and one fungus (Candida glabrata) were tested once; Escherichia coli was tested twice. As shown in the left section of Table 1, both BD PureHub™ and ICU Medical™ tested negative. All the NFCs cultured were positive to confirm the correct contamination process.

Table 1

Basal Test and Stress Test performed with different pathogens (column 1) on the 2 types (BD PureHub™ and ICU Medical™) of port protectors. NFC means Needle free Connectors. 90 and 180 minutes are times after microbial incubation. N means negative

BASAL TEST BD PureHub™ + NFC ICU Medical™ + NFC STRESS TEST BD PureHub™ ICU Medical™
90 180 90 180 90 180 90 180
Carbapenem-resistant
Klebsiella pneumoniae
NNNNGlycopeptide-resistant
Enterococcus faecalis 1
NNNN
Multidrug-resistant
Pseudomonas aeruginosa
NNNNGlycopeptide-resistant
Enterococcus faecalis 2
NNNN
Escherichia coli 1NNNNMultidrug-resistant
Pseudomonas aeruginosa 1
NNNN
Escherichia coli 2NNNNmultidrug-resistant
Pseudomonas aeruginosa 2
NNNN
Methicillin-resistant
Staphylococcus aureus
NNNNMultidrug-resistant
Pseudomonas aeruginosa 3
NNNN
Multidrug-resistant
Acinetobacter pittii
NNNNMethicillin-resistant
Staphylococcus aureus 1
NNNN
Candida glabrataNNNNMethicillin-resistant
Staphylococcus aureus 2
NNNN
Methicillin-resistant
Staphylococcus aureus 3
NNNN
Escherichia coli 1NNNN
Escherichia coli 2NNNN
Streptococcus agalactiae 1NNNN
Streptococcus agalactiae 2NNNN
Streptococcus agalactiae 3NNNN
Serratia marcescens 1NNNN
Serratia marcescens 2NNNN
Serratia marcescens 3NN10^410^4
Corynebacterium striatum 1NNNN
Corynebacterium striatum 2NNNN
Corynebacterium striatum 3NNNN
Acinetobacter baumannii MDR 110^310^3NN
Acinetobacter baumannii MDR 2NNNN
Acinetobacter baumannii MDR 3NNNN
Candida glabrataNNNN
Staphylococcus capitis 1NNNN
Staphylococcus capitis 2NNNN
Staphylococcus capitis 3NNNN
Staphylococcus epidermidis 1NNNN
Staphylococcus epidermidis 2NNNN
Staphylococcus epidermidis 3NNNN

As shown in the right section of Table 1, fifty-eight assays were performed in the stress test: 3 Gram negative (multidrug-resistant P. aeruginosa, Serratia marcescens, MDR Acinetobacter baumannii), and several Gram positive (methicillin-resistant S. aureus, Streptococcus agalactiae, Corynebacterium striatum, Staphylococcus capitis and Staphylococcus epidermidis) were tested three times; glycopeptide-resistant Enterococcus faecalis and E. coli twice; C. glabrata was tested once.

The only significant bacterial load was found for S. marcescens (104 CFU/mL in ICU Medical™), both at 90 and 180 minutes after incubation; due to the low microbial load, MDR Acinetobacter baumannii was not considered significant (<103 CFU/mL in BD PureHub™). All the cultured catheters hub tested positive to confirm the correct contamination process.

DISCUSSION

There are no in vitro and in vivo studies comparing DCs. Moreover, no information about the sponge dimension and the amount of 70% isopropyl alcohol are displayed in the datasheets [19, 20]. Before the beginning of the study, we have asked the manufacturers for the above information and they answered that those data were not public; thus, we remained blind to some important structural differences that could have been useful in DCs comparative evaluation.

To the best of our knowledge, this is the first independent study that evaluates DCs antimicrobial effectiveness and that brings out potential differences between what manufacturers have claimed in their datasheets.

With the aim to compare the results on microbial strains claimed by manufacturers, our study evaluated, in an in vitro model, the ability to suppress bacterial growth of two of the many commercially available DCs: BD PureHub™ and ICU Medical™, despite micro-organisms resistance patterns.

In the in vitro model we chose a high load of bacterial contamination, in order to reproduce the worst clinical settings. The use of femoral catheters, mainly with mid-thigh exit site, is becoming increasingly common in hospitals, particularly in bedridden patients with psychomotor agitation, delirium, and dementia [2122]. These patients are, among others, those at greater risk of bacterial skin colonization; indeed, many of them live in nursing homes, have had several hospital admissions and undergo bed hygiene [21]. During this cleaning procedure, it is common to get the catheter’s hub dirty and contaminated with a high bacterial load (>0.5 McFarland). The microbiological conditions in this clinical setting could be considered similar to the ones recreated in the present study.

Another clinical setting to be mentioned is the presence of a catheter hub close to a tracheostomy, where respiratory secretions are abundant. During the cleaning of the stoma, even in well positioned venous catheters (with exit site at least 10 cm away from the stoma), a high microbial contamination could happen when aspiration of secretions is performed.

In the baseline test, no significant microbial growth was observed, confirming that the use of NFCs in combination with Port Protectors significantly reduces the risk of CRBSIs [17]. Thus, respecting the aseptic technique of accessing to the system, NFCs are difficult to contaminate.

In the stress test the results obtained in vitro for both DCs were promising; only in one out of fifty-eight trials the bacterial growth was significant. Therefore, our results confirm what has been reported in the BD PureHub™ datasheet and add data not previously shown by ICU Medical™ [19, 20].

Moreover, no difference was shown between the two manufacturers products: the use DCs on NFCs was able to reclaim the catheter lumen in situations. These findings support the routine use of DCs together with NFCs to reduce the incidence of CRBSIs, as part of a structured bundle of interventions.

Finally, the stress test confirms what the producers claim: DCs cannot be used merely as disinfectants of the distal hub, on which the manual scrub-the-hub technique is needed.

A special consideration should be made about the two positive results on S. marcescens and A. baumannii cultures: both are gram-negative, hospital-acquired pathogens responsible for several nosocomial infections, especially CRBSIs [23]. The two bacteria are well known producers of biofilm and that means an advantage on catheters colonization process [24]. Bacterial biofilm is a complex surface-adhered community of viable and dead bacteria, encased within an extracellular matrix composed of polysaccharide, proteins and extracellular DNA. Bacteria in biofilm also exhibit increased resistance to antimicrobial agents for reasons such as resistance to drug penetration and to desiccation, due to metabolic changes such as slower growth rates [25]. Even if biofilm production could be one of the major explanations, other virulence factors cannot be excluded; for this reason, further investigations are needed.

This study has some potential limitations: first of all, we analyzed only one fungus, which was tested once in both basal and stress test; furthermore, this is a monocentric study so it was not possible to highlight any differences in the results among different centers.

As suggested by the recent SHEA guidelines 2022, a single recommendation could be ineffective when used alone: DCs are useful as part of a broad set of strategies to prevent CRBSIs [26].

Further investigations are needed to confirm our data both in vitro and in clinical settings and to compare other manufacturers products.

Footnotes

Conflict of interest: The Authors declare no conflicts of interest.

Funding: All the material used in this in vitro study were provided by our University Hospital ASST Fatebenefratelli-Sacco Milan, Italy, in order to evaluate the best disinfectant cap available on the market.

Avalaibility of data

The data supporting the findings of this study are available from the corresponding author upon reasonable request.

REFERENCES

1. Chaves F, Garnacho-Montero J, Del Pozo JL, et al. Diagnosis and treatment of catheter-related bloodstream infection: Clinical guidelines of the Spanish Society of Infectious Diseases and Clinical Microbiology and (SEIMC) and the Spanish Society of Spanish Society of Intensive and Critical Care Medicine and Coronary Units (SEMICYUC) Med intensiva (Engl Ed) 2018;42(1):5–36. [Abstract] [Google Scholar]
2. Vaz LE, Kleinman KP, Kawai AT, et al. Impact of Medicare’s hospital-acquired condition policy on infections in safety net and non–safety net hospitals. Infect Control Hosp Epidemiol. 2015;36(6):649–655. [Abstract] [Google Scholar]
3. Gidaro A, Vailati D, Gemma M, et al. Retrospective survey from vascular access team Lombardy net in COVID-19 era. J Vasc Access. 2022;23(4):532–537. [Abstract] [Google Scholar]
4. Centers for Disease Control and Prevention. National and State Healthcare-Associated Infections Progress Report. 2014. Published March 2016. Available at www.cdc.gov/hai/progress-report/index.html.
6. Rosenthal VD. Clinical impact of needle-free connector design: A systematic review of literature. J Vasc Access. 2020;21(6):847–853. [Abstract] [Google Scholar]
7. Ture Z, Blot S, Alp E. Central line-associated bloodstream infection prevention: “scrub the hub” or antiseptic barrier caps? Intensive Crit Care Nurs. 2023;77:103442. [Abstract] [Google Scholar]
8. Marilyn Hanchett RN.Needleless connectors and bacteremia: is there a relationship. Infect Control Today. 2005. Avaible at: https://www.infectioncontroltoday.com/view/needleless-connectors-and-bacteremia-there-relationship.
9. Menyhay SZ, Maki DG. Disinfection of needleless catheter connectors and access ports with alcohol may not prevent microbial entry: the promise of a novel antiseptic-barrier cap. Infect Control Hosp Epidemiol. 2006;27(1):23–27. [Abstract] [Google Scholar]
10. Oto J, Imanaka H, Konno M, Nakataki E, Nishimura M. A prospective clinical trial on prevention of catheter contamination using the hub protection cap for needleless injection device. Am J Infect Control. 2011;39(4):309–313. [Abstract] [Google Scholar]
11. Inchingolo R, Pasciuto G, Magnini D, et al. Educational interventions alone and combined with port protector reduce the rate of central venous catheter infection and colonization in respiratory semi-intensive care unit. BMC Infect Dis. 2019;19(1):215. [Europe PMC free article] [Abstract] [Google Scholar]
12. Martino A, Thompson L, Mitchell C, et al. Efforts of a Unit Practice Council to implement practice change utilizing alcohol impregnated port protectors in a burn ICU. Burns. 2017;43(5):956–964. [Abstract] [Google Scholar]
13. Ramirez C, Lee AM, Welch K. Central venous catheter protective connector caps reduce intraluminal catheter-related infection. J Assoc Vasc Access. 2012;4(17):210–213. [Google Scholar]
14. Merrill KC, Sumner S, Linford L, Taylor C, Macintosh C, et al. Impact of universal disinfectant cap implementation on central line–associated bloodstream infections. Am J Infect Control. 2014;42(12):1274–1277. [Abstract] [Google Scholar]
15. Gillis VELM, van Es MJ, Wouters Y, Wanten GJA. Antiseptic barrier caps to prevent central line-associated bloodstream infections: A systematic review and meta-analysis. Am Journal Infect Control. 2023;51(7):827–835. [Abstract] [Google Scholar]
16. Loveday HP, Wilson JA, Prieto J, Wilcox MH. epic3: revised recommendation for intravenous catheter and catheter site care. J Hosp Infect. 2016;92(4):346–348. [Abstract] [Google Scholar]
17. Barton A. The case for using a disinfecting cap for needlefree connectors. Br J Nurs. 2019;28(14):S22–S27. [Abstract] [Google Scholar]
18. Opota O, Croxatto A, Prod’hom G, Greub G. Blood culture-based diagnosis of bacteraemia: state of the art. Clin Microbiol Infect. 2015;21(4):313–22. [Abstract] [Google Scholar]
21. Bartoli A, Donadoni M, Quici M, et al. Safety of mid-thigh exit site venous catheters in multidrug resistant colonized patients. J Vasc Access. 2023:11297298231188150. [Abstract] [Google Scholar]
22. Gidaro A, Samartin F, Salvi E, et al. Midline peripheral catheters inserted in the superficial femoral vein at mid-thigh: Wise choice in COVID-19 acute hypoxemic respiratory failure patients with helmet continuous positive airway pressure. J Vasc Access. 2022:11297298221085450. [Abstract] [Google Scholar]
23. Bukhari S, Banjar A, Baghdadi S, Baltow B, Ashshi A, Hussain W. Central line associated blood stream infection rate after intervention and comparing outcome with national healthcare safety network and international nosocomial infection control consortium data. Ann Med Health Sci Res. 2014;4(5):682–686. [Europe PMC free article] [Abstract] [Google Scholar]
24. Satpathy S, Sen SK, Pattanaik S, Raut S. Review on bacterial biofilm: an universal cause of contamination. Biocatal Agric Biotechnol. 2016;7:56–66. [Google Scholar]
25. Choudhary M, Shrivastava R, Vashistt J. Acinetobacter baumannii Biofilm Formation: Association with Antimicrobial Resistance and Prolonged Survival under Desiccation. Curr Microbiol. 2022;79(12):361. [Abstract] [Google Scholar]
26. Buetti N, Marschall J, Drees M, et al. Strategies to prevent central line-associated bloodstream infections in acute-care hospitals: 2022 Update. Infect Control Hosp Epidemiol. 2022;43(5):553–569. [Europe PMC free article] [Abstract] [Google Scholar]

Articles from Le Infezioni in Medicina are provided here courtesy of Edizioni Internazionali s.r.l.

Citations & impact 


This article has not been cited yet.

Impact metrics

Alternative metrics

Altmetric item for https://www.altmetric.com/details/157383550
Altmetric
Discover the attention surrounding your research
https://www.altmetric.com/details/157383550